5. Remove the main motor-fan [11] including the side stirrup [12].<br />
12<br />
6. Position the new motor-fan after having mounted again the flow sensor group [15] and the relevant<br />
stirrup.<br />
7. Fasten the motor-fan with the relevant screws and reconnect electrical connectors.<br />
8. Proceed as indicated in the sub section "Replacement of HEPA filters", from 13 to 15 points.<br />
Replacement of exhaust motor-fan (only for <strong>CytoFAST</strong> <strong>Elite</strong>).<br />
1. Proceed as indicated in sub-section "Replacement of HEPA filters", from 1 to 4 points and remove the<br />
exhaust HEPA filter.<br />
2. Remove the textile plenum<br />
3. Remove the anemoetric fan [14]<br />
<strong>00214EN</strong> <strong>Rev.05</strong> – 12/2011 33<br />
11