a D:GROUP company<br />
3 OPERATION PRINCIPLES<br />
The following are the working principles of the CYTOFAST ELITE-cabinet :<br />
the pressurized air pushed into the plenum of the main motor-fan passes through the absolute filter and then<br />
downwards, in laminar flow, into the working chamber (A). From here, through the slots of the work surface,<br />
having mixed with the external air (B) which enters the cabinet from the front opening, the air passes through<br />
the absolute filter placed under the work surface (C) and is sucked into the intake up channel situated at the<br />
rear of the working area (D). Part of the air (E) is extracted by the second motor-fan through the<br />
corresponding absolute exhaust filter. This extracted air generates the intake of air from the outside (front<br />
barrier) which ensures the protection of the operator/environment from hazardous material manipulated in the<br />
working area.<br />
The extracted air may be filtered again through a second absolute exhaust filter placed on the exhaust duct of<br />
the cabinet (BS 5726), if the air is recycled inside the laboratory where the cabinet is installed.<br />
E<br />
A<br />
C<br />
<strong>00214EN</strong> <strong>Rev.05</strong> – 12/2011 7<br />
D<br />
C<br />
E<br />
A<br />
B