05.08.2013 Views

00214EN Rev.05 - CytoFAST Elite - Wolf Laboratories

00214EN Rev.05 - CytoFAST Elite - Wolf Laboratories

00214EN Rev.05 - CytoFAST Elite - Wolf Laboratories

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

a D:GROUP company<br />

3 OPERATION PRINCIPLES<br />

The following are the working principles of the CYTOFAST ELITE-cabinet :<br />

the pressurized air pushed into the plenum of the main motor-fan passes through the absolute filter and then<br />

downwards, in laminar flow, into the working chamber (A). From here, through the slots of the work surface,<br />

having mixed with the external air (B) which enters the cabinet from the front opening, the air passes through<br />

the absolute filter placed under the work surface (C) and is sucked into the intake up channel situated at the<br />

rear of the working area (D). Part of the air (E) is extracted by the second motor-fan through the<br />

corresponding absolute exhaust filter. This extracted air generates the intake of air from the outside (front<br />

barrier) which ensures the protection of the operator/environment from hazardous material manipulated in the<br />

working area.<br />

The extracted air may be filtered again through a second absolute exhaust filter placed on the exhaust duct of<br />

the cabinet (BS 5726), if the air is recycled inside the laboratory where the cabinet is installed.<br />

E<br />

A<br />

C<br />

<strong>00214EN</strong> <strong>Rev.05</strong> – 12/2011 7<br />

D<br />

C<br />

E<br />

A<br />

B

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!