Post-fumigation cleaning<br />
Having performed fumigation, the manipulation of the filters is safe: however, this manipulation should be<br />
carried out by technicians wearing protective gloves and clothing.<br />
Before using the cabinet again any possible remains of the foregoing procedure of vaporization and relevant<br />
neutralization (which can show, for example, as white dust) must be removed from the work chamber.<br />
ATTENTION: for cabinets that are not connected to the outside, but recycling the air in the<br />
environment, make sure that ambient concentration does not exceed 2.5mg/m 3 , as<br />
higher concentrations are harmful.<br />
Faster S.r.l. refuses all responsibility for the performance of such sterilization procedures or of any<br />
possible consequences deriving from it, reminding that the full responsibility for the operation lies<br />
only and exclusively with the properly trained operator-user.<br />
7.D REPLACEMENT OF HEPA FILTERS (by technical assistance personnel)<br />
Important: before replacing HEPA filters, it is absolutely necessary that:<br />
- if pathogens have been manipulated, the cabinet must be decontaminated and the user must<br />
issue a sterilization certificate to the personnel before the start of the operation.<br />
- for the safety of the personnel and the environment, the use of PVC gloves is recommended as<br />
well as the collection of the replaced HEPA filters in polyethylene bags.<br />
Replacement of MAIN/EXHAUST HEPA filter (see diagram for maintenance operations)<br />