Post-fumigation cleaning<br />

Having performed fumigation, the manipulation of the filters is safe: however, this manipulation should be<br />

carried out by technicians wearing protective gloves and clothing.<br />

Before using the cabinet again any possible remains of the foregoing procedure of vaporization and relevant<br />

neutralization (which can show, for example, as white dust) must be removed from the work chamber.<br />

ATTENTION: for cabinets that are not connected to the outside, but recycling the air in the<br />

environment, make sure that ambient concentration does not exceed 2.5mg/m 3 , as<br />

higher concentrations are harmful.<br />

Faster S.r.l. refuses all responsibility for the performance of such sterilization procedures or of any<br />

possible consequences deriving from it, reminding that the full responsibility for the operation lies<br />

only and exclusively with the properly trained operator-user.<br />

7.D REPLACEMENT OF HEPA FILTERS (by technical assistance personnel)<br />

Important: before replacing HEPA filters, it is absolutely necessary that:<br />

- if pathogens have been manipulated, the cabinet must be decontaminated and the user must<br />

issue a sterilization certificate to the personnel before the start of the operation.<br />

- for the safety of the personnel and the environment, the use of PVC gloves is recommended as<br />

well as the collection of the replaced HEPA filters in polyethylene bags.<br />

Replacement of MAIN/EXHAUST HEPA filter (see diagram for maintenance operations)<br />

<strong>00214EN</strong> <strong>Rev.05</strong> – 12/2011 27