8. Cleaning and maintenance<br />
Universal cassette clamp<br />
54<br />
60<br />
13<br />
Fig. 57<br />
8.2 Maintenance instructions<br />
Detach cassette clamp (13) for a thorough cleaning, removing all<br />
paraffin residues.<br />
For cleaning, do not use xylene. Use xylene substitutes or paraffin<br />
removers such as "Para Gard."<br />
The cassette clamp (13) can also be placed in an oven heated to a<br />
maximum <strong>of</strong> 65 °C, until the liquid wax escapes.<br />
Remove paraffin residues with a dry cloth.<br />
After such a cleaning procedure in an oven, always be sure to lubricate<br />
the axle and spring <strong>of</strong> the clamping lever (60) (see also chapter 8.3).<br />
Only authorized and qualified service personnel may access the internal components <strong>of</strong> the instrument<br />
for service and repair!<br />
The instrument is basically maintenance-free. To ensure trouble-free operation <strong>of</strong> the instrument<br />
over a long period <strong>of</strong> time, the following is recommended by Leica:<br />
1. Thoroughly clean the instrument on a daily<br />
basis.<br />
2. Enter into a service contract at the latest at<br />
the end <strong>of</strong> the warranty period. For more information,<br />
please contact your local Leica<br />
technical service center.<br />
3. Have the instrument inspected on a regular<br />
basis by a qualified service technician<br />
authorized by Leica.<br />
The intervals depend on how heavily the<br />
instrument is used.<br />
We recommend the following inspection intervals depending on the general workload <strong>of</strong> the instrument<br />
as defined in the two categories listed in the table:<br />
Category I<br />
Category II<br />
Number <strong>of</strong> sections per day: > 8.000 sections per day<br />
< 8.000 sections per day<br />
Workload (hours per day): > 5 hours per day<br />
< 5 hours per day<br />
Sectioning speed:<br />
Predominantly high sectioning speed Slow to medium sectioning speed<br />
Specimen material:<br />
Working with s<strong>of</strong>t and hard specimens Predominantly s<strong>of</strong>t specimen material<br />
Maintenance:<br />
Every 12 months<br />
Every 24 months<br />
Instruction manual V 1.4 – 02/2010