proposed dual sample. The dual sample affords greater precision than the single sample due to its<br />

greater effective thickness of 36 mm.<br />

5.2 EXPERIMENTAL<br />

The experiment was performed at the S18 beam port of 58 MW Institut Laue Langevin (ILL)<br />

reactor in Grenoble, France. The Si perfect-crystal monochromator and IFM are operated in a nondispersive<br />

configuration which provides a narrow rocking curve with 1.8" FWHM at a wavelength<br />

of 1.8Å. To attain greater contrast and phase stability, the optical bench is shielded against<br />

Fig.57 A photograph of the experimental setup at S18 ILL, depicting a pair of amorphous Si prisms<br />

each of 120 o apex angles placed before the symmetric {220} LLL IFM to remove the λ/2 from<br />

monochromatic neutron beam. Dual sample mounted on a plate, is hanging from the top.<br />

106