MALDI-TOF MS Target Preparation<br />

MALDI-TOF MS target preparation is one of the most critical steps <strong>for</strong> high quality and reproducible<br />

results. Depending on the application, different target types and MALDI matrix preparations are<br />

favored. For all applications it is mandatory to use a thoroughly cleaned target plate. In<br />

dependence on the type of target <strong>Bruker</strong> Daltonic recommends different cleaning procedures.<br />

Cleaning of MALDI-MS targets<br />

Cleaning of AnchorChip TM targets<br />

This is an advanced cleaning protocol differing from the standard cleaning procedure<br />

recommended in the product description.<br />

1. Rinse the target intensively under<br />

flowing hot tap water<br />

Hot tap water<br />

2. Wipe the target intensively with acetone<br />

using a kimwipe kimwipe.<br />

: Acetone is highly flammable and<br />

irritant (R: 11-36-66-67; S: (2)-9-16-26)<br />

Acetone<br />

3. Rinse the target with distilled water<br />

(Milli-Q).<br />

Distilled water<br />

4. Rinse the target with methanol and let it<br />

dry.<br />

: Methanol is highly flammable and<br />

toxic (R: 11-23/24/25-39/23/24/25; S:<br />

(1/2)-7-16-36/37-45)<br />

Methanol<br />