Instructions for Use - Glyco Kit MB-LAC Con A - Bruker
Instructions for Use - Glyco Kit MB-LAC Con A - Bruker
Instructions for Use - Glyco Kit MB-LAC Con A - Bruker
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
MALDI-TOF MS Target Preparation<br />
MALDI-TOF MS target preparation is one of the most critical steps <strong>for</strong> high quality and reproducible<br />
results. Depending on the application, different target types and MALDI matrix preparations are<br />
favored. For all applications it is mandatory to use a thoroughly cleaned target plate. In<br />
dependence on the type of target <strong>Bruker</strong> Daltonic recommends different cleaning procedures.<br />
Cleaning of MALDI-MS targets<br />
Cleaning of AnchorChip TM targets<br />
This is an advanced cleaning protocol differing from the standard cleaning procedure<br />
recommended in the product description.<br />
1. Rinse the target intensively under<br />
flowing hot tap water<br />
Hot tap water<br />
2. Wipe the target intensively with acetone<br />
using a kimwipe kimwipe.<br />
: Acetone is highly flammable and<br />
irritant (R: 11-36-66-67; S: (2)-9-16-26)<br />
Acetone<br />
3. Rinse the target with distilled water<br />
(Milli-Q).<br />
Distilled water<br />
4. Rinse the target with methanol and let it<br />
dry.<br />
: Methanol is highly flammable and<br />
toxic (R: 11-23/24/25-39/23/24/25; S:<br />
(1/2)-7-16-36/37-45)<br />
Methanol<br />