Cleaning of polished steel targets<br />
This is an advanced cleaning protocol differing from the standard cleaning procedure<br />
recommended in the respective target’s product description.<br />
1. Rinse the target intensively under<br />
flowing hot tap water.<br />
Hot tap water<br />
2. Wipe the target intensively with acetone<br />
using a kimwipe.<br />
: Acetone is highly flammable and<br />
irritant (R: 11-36-66-67; S: (2)-9-16-26)<br />
Acetone<br />
3. Rinse the target intensively under<br />
flowing hot tap water.<br />
Hot tap water<br />
4. Wipe the target intensively with 80%<br />
TFA using a kimwipe.<br />
: 80% TFA is corrosive (R: 20-35-<br />
52/53; S: (1/2)-9-26-27-28-45-61)<br />
80 % TFA<br />
5. Rinse the target intensively under<br />
flowing hot tap water.<br />
Hot tap water<br />
6. Rinse the target with distilled water<br />
(Milli-Q).<br />
Distilled water<br />
7. Rinse the target with methanol and let it<br />
dry.<br />
: Methanol is highly flammable and<br />
toxic (R: 11-23/24/25-39/23/24/25; S:<br />
(1/2)-7-16-36/37-45)<br />
Methanol<br />