Cleaning of polished steel targets<br />

This is an advanced cleaning protocol differing from the standard cleaning procedure<br />

recommended in the respective target’s product description.<br />

1. Rinse the target intensively under<br />

flowing hot tap water.<br />

Hot tap water<br />

2. Wipe the target intensively with acetone<br />

using a kimwipe.<br />

: Acetone is highly flammable and<br />

irritant (R: 11-36-66-67; S: (2)-9-16-26)<br />

Acetone<br />

3. Rinse the target intensively under<br />

flowing hot tap water.<br />

Hot tap water<br />

4. Wipe the target intensively with 80%<br />

TFA using a kimwipe.<br />

: 80% TFA is corrosive (R: 20-35-<br />

52/53; S: (1/2)-9-26-27-28-45-61)<br />

80 % TFA<br />

5. Rinse the target intensively under<br />

flowing hot tap water.<br />

Hot tap water<br />

6. Rinse the target with distilled water<br />

(Milli-Q).<br />

Distilled water<br />

7. Rinse the target with methanol and let it<br />

dry.<br />

: Methanol is highly flammable and<br />

toxic (R: 11-23/24/25-39/23/24/25; S:<br />

(1/2)-7-16-36/37-45)<br />

Methanol<br />