Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Before starting the functional test, the hand probe must be switched<br />

on for approx. 5 min. at a minimum and run in "clean" surrounding air<br />

(minimal or no background concentration of the gas types that are<br />

specific to the special type). These conditions are required for calculation<br />

of the zero point. If no zero point can be set, the functional test<br />

is stopped. The device must then be restarted and the functional test<br />

must be repeated after waiting 5 min.<br />

Continuous contact during the functional test must be ensured between<br />

the Teflon tip of the hand probe with the head of the test leakage.<br />

Only manufacturer-approved reference leaks can be used for the<br />

functional test.<br />

The functional test must be carried out in enclosed spaces with minimal<br />

air movement.<br />