Before starting the functional test, the hand probe must be switched<br />
on for approx. 5 min. at a minimum and run in "clean" surrounding air<br />
(minimal or no background concentration of the gas types that are<br />
specific to the special type). These conditions are required for calculation<br />
of the zero point. If no zero point can be set, the functional test<br />
is stopped. The device must then be restarted and the functional test<br />
must be repeated after waiting 5 min.<br />
Continuous contact during the functional test must be ensured between<br />
the Teflon tip of the hand probe with the head of the test leakage.<br />
Only manufacturer-approved reference leaks can be used for the<br />
functional test.<br />
The functional test must be carried out in enclosed spaces with minimal<br />
air movement.<br />