Multi Gas Leak Detector <strong>MGLD.3000</strong><br />
Operating instructions<br />
Calibration procedure<br />
Warnings / functions<br />
Warning symbol - blinks<br />
during check and calibration<br />
Probe is initialized<br />
The sensor tip must be held at the reference leak within 5 seconds. If the<br />
reference leak is detected (through an increase in the concentration) the functional<br />
test begins, otherwise the procedure is canceled. Make sure the reference<br />
leak is opened and is filled with test gas in the specified pressure range.<br />
The functional test must be repeated.<br />
If the measured concentration value is inside the specified range (set-point +/-<br />
30%), the functional test is completed successfully and a correction factor is<br />
set to adjust the sensitivity.<br />
Cancel and return to the previous leak search menu<br />
Tracing<br />
Warnings / functions<br />
Warning symbol - blinks<br />
during check and calibration<br />
Probe is initialized<br />
The sensitivity level of the hand probe is indicated by the position of the<br />
pointer in the pointer instrument. If the sensitivity matches the set-point values,<br />
this is indicated by a pointer position in the middle green range of the<br />
scale. The functional test was passed. If the sensitivity is within the specified<br />
tolerance limits, but is smaller or higher than the set-point, this is indicated by<br />
a pointer that is in the left or right yellow part of the scale range. The functional<br />
test was passed and a correction factor was calculated.<br />
BAL_MGLD3000_20101001_en.doc 33-54