Multi Gas Leak Detector <strong>MGLD.3000</strong><br />
Operating instructions<br />
6212, 6213, Probe: Sensor<br />
6216, 6217, Probe: Calibration Data<br />
6218, 6221, Probe: Identification<br />
6222, 6223,<br />
6224<br />
6300 Base: No Communication<br />
6301 Base: Reduced Flow<br />
6302 Base: No Flow<br />
The device does not turn on.<br />
These error numbers indicate a defect in the gas sensor system<br />
or a storage error in the hand probe. If the error occurs again<br />
after restarting the device, contact technical support.<br />
There is a problem with the internal device communication between<br />
the individual components. If the error occurs again after<br />
restarting the device, contact technical support.<br />
The current in the base unit is monitored by a current sensor. A<br />
current that is too low, or a complete absence of current, in the<br />
main device can have the following causes:<br />
• Contamination/clogging of the Teflon tip in the hand probe<br />
• Damage (kinks in the connection line)<br />
• Closed inline filter in the main device<br />
After correcting the root problem, the current is re-established.<br />
The determination of the current can take up to 30 sec.<br />
In case of error, the measurement characteristics of the device<br />
can be reduced. A determination of the gas concentration is still<br />
possible, although the conversion to leak rates is incorrect.<br />
The battery is dead and the device is not connected to the power<br />
supply.<br />
Check for a proper power connection. Check for proper positioning<br />
on the docking unit.<br />
BAL_MGLD3000_20101001_en.doc 53-54