5. Care and service<br />
Please follow the general safety instructions in 2.2 of these operating instructions.<br />
Interventions in the device, service or repair works can only be performed by the manufacturer's customer<br />
service department or by authorized professional technicians.<br />
5.1. Cleaning and care<br />
Remove the power plugs from the socket before cleaning.<br />
External cleaning of the <strong>MGLD.3000</strong> and its components may only be performed using a soft,<br />
lightly moistened, non-dripping cloth. Chemical cleaning agents, abrasives, or solvents may not be used.<br />
The contacts between the docking unit and the base unit, between the device and connection line,<br />
probe and connection line, and the interfaces with the rear of the device may only be cleaned with a dry,<br />
dust-free brush.<br />
During longer periods between use, the device and hand probes must be wrapped in their original packaging<br />
materials and placed in storage. Especially with probes, the storage area must be absolutely free of<br />
any impurities and contaminants, and must be specially protected from the influence of chemical substances,<br />
e.g. silicone.<br />
5.2. Service<br />
Before startup, regularly check the electro-pneumatic connection line for kinks, crushing or compression<br />
damage, cracks and other damage and/or impurities. Replace damaged connection lines immediately to<br />
ensure proper leak search results and to protect the device from contamination.<br />
Also regularly check the Teflon tip and the ventilation inlets/outlets on the rear of the device before startup<br />
for dirt buildup and contamination and remove these.<br />
Generally a yearly check of the hand probes by the manufacturer's customer service or by manufacturerauthorized<br />
technicians / an authorized service center is recommended.<br />
We also recommend yearly replacement of all filters by the manufacturer customer service department or<br />
by manufacturer-authorized specialists / an authorized service center.<br />