Multi Gas Leak Detector <strong>MGLD.3000</strong><br />
Operating instructions<br />
The following error messages can be output:<br />
Error code Error message Cause / recommendations<br />
6101 No Probe present No probe was connected to the base unit.<br />
Switch off the device. Check the connection on the electropneumatic<br />
connection line on the base unit and on the probe.<br />
Check the connection line for damage such as kinks or cracks.<br />
Switch the device back on.<br />
6201 Probe: Heating Sensor The hand probe is initialized and the sensor is heated. This warning<br />
message will disappear when the sensor has finished heating<br />
(approx. 1-2 min.). If the warning message is still displayed after<br />
2 min., restart the device.<br />
6202 Probe: Reading Calibration<br />
Data<br />
The hand probe is initialized and the calibration data is read. This<br />
message shows an error while reading the calibration data from<br />
the probe. To correct the error, restart the device.<br />
6203 Probe: Writing Calibration<br />
Data<br />
Once the test- and calibration procedure is completed, the correction<br />
factor that is calculated is saved in the probe. The message<br />
indicates that an error occurred while saving the correction<br />
factor. To ensure correct transfer of the correction factor, repeat<br />
the test- and calibration procedure.<br />
6204 Probe: Reduced Flow The current through the probe is monitored by a current sensor.<br />
6205 Probe: No Flow A current that is too low, or a complete absence of current, in the<br />
hand probe can have the following causes:<br />
• Contamination/clogging of the Teflon tip<br />
• Damage (cracks, kinks in the connection line)<br />
• Insufficient connection of the probe to the connection line<br />
After correcting the root problem, the current is re-established. It<br />
can take up to 30 sec. to determine the current.<br />
In case of error the measurement characteristics of the device<br />
can be reduced. A determination of the gas concentration is still<br />
possible, although the conversion to leak rates is incorrect.<br />
BAL_MGLD3000_20101001_en.doc 52-54