Ryan’s MFP-3D Procedural Operation ‘Manualette’ Version 10 (v080501; Igor 6.03A); ©2008 6.16<br />

A<br />

B<br />

C<br />

D<br />

Figure 6.1.12: Q gain and repulsive imaging: A) post auto tune- no Q gain; B)almost ringing; C) ringing; D)<br />

freakout.<br />

If all else fails, the tip could be too blunt (recall larger surface areas promote attractive mode- see Table 6.1.4)-<br />

CHANGE THE TIP; or the sample could possess charge that needs to be quelled/ dissipated using a ‘<strong>St</strong>atic<br />

Master’ (see Figure 6.1.12).<br />

TROUBLESHOOTING-<br />

If you are still having trouble with mode hopping, it could be due to the sample - here are a couple things that<br />

may be contributing to it-<br />

1) If you are experiencing difficulty keeping the tip in repulsive mode, perhaps the tip is too dull (see Table<br />

6.1). If you are re-using the tip, this is a strong likely hood.<br />

2) Surface charges- use the <strong>St</strong>atic Master device to dissipate the charge. Place the <strong>St</strong>atic Master in the vicinity<br />

of the sample (see Figure 6.1.13). Sometimes the glass slide the sample is glued to is the culprit of the excess<br />

charge- in these cases, mounting the sample on a magnetic puck and placing it on a metal sample holder can<br />

help quell this charge.<br />

<strong>St</strong>atic master<br />

Figure 6.1.13: Placing static master near sample to dissipate charge.