CHAPTER 3<br />
QC AND QA LABORATORIES<br />
3.1 General. All laboratories involved in the design, design verification, quality control,<br />
and quality assurance testing of asphalt mixtures and component materials must be<br />
fully equipped to perform the required tests. All such equipment must meet the<br />
requirements of MDOT’s specified test methods.<br />
3.2 QC Laboratory. Post an organization chart, including names, telephone numbers<br />
and current certification (CAT-I, CAT-II and CMDT), of all personnel responsible<br />
for the quality control program on the project. Post the chart in a readily visible<br />
location.<br />
Post all required quality control charts in a readily visible location, and maintain in<br />
an up-to-date status. Charts may be maintained on a computer. However, updated<br />
computer charts shall be printed and displayed at a minimum of once each<br />
production day.<br />
The laboratory shall be setup and calibrated prior to production. Calibration shall be<br />
according to the procedures and frequencies given in AASHTO R 18. Inspect and<br />
calibrate the laboratory equipment as follows:<br />
Check calibration of Gyratory compactor according to the requirements in the<br />
manufacturer’s manual (at a minimum, verify the angle of gyration and the load<br />
on the specimen during compaction).<br />
Check balances for accuracy, precision and readability in accordance with<br />
AASHTO M 231.<br />
Check the water baths used for specific gravity testing for constant water level,<br />
temperature, and make sure the specimen basket does not touch the walls of the<br />
tank each time testing is performed. Make sure a non-corrosive metal wire or a<br />
non-absorptive cord is used to hang the basket from the bottom of the balance.<br />
Check calibration of maximum specific gravity flask and fine aggregate<br />
angularity cylinder.<br />
Check vacuum pumps to confirm that the required pressure is being applied to<br />
the flask or pycnometer.<br />
Check accuracy of all thermometers.<br />
Check temperature of ovens and water baths.<br />
Rev. February 5, 2007 12