20 Getting started <strong>Welch</strong> <strong>Allyn</strong> <strong>Propaq</strong> <strong>Encore</strong> <strong>Vital</strong> <strong>Signs</strong> <strong>Monitor</strong><br />

<strong>Monitor</strong> setup<br />

Setup Menu 1 is accessed by pressing the SETUP button on the Main Menu.<br />

STATSCALE<br />

ALARMS<br />

WAVE SEL<br />

TRENDS<br />

MORE<br />

STATSCALE<br />

ALARMS<br />

WAVE SEL<br />

TRENDS<br />

MORE<br />

NEXT<br />

CHANGE<br />

PRINTER<br />

MORE<br />

CURRENT<br />

SOURCE<br />

SELECTED<br />

SOURCE<br />

SWEEP (mm/s)<br />

ALARM TONE<br />

Automatically readjusts all waveform scales.<br />

Allows access to the Alarms menu.<br />

Allows you to turn on and off desired waveforms or NIBP numerics<br />

for display.<br />

Allows access to the Trend settings and display.<br />

Displays the next setup menu and the following status window:<br />

SETUP<br />

HR/PR RR/BR ( 85)<br />

CURRENT SOURCE : ECG MCO2 PA 35/ 18<br />

SELECTED SOURCE : ART --- MCO2 BR<br />

SWEEP (mm/s) : 12.5 3.13<br />

ALARM TONE : HIGH<br />

35<br />

mmHg<br />

HR/PR TONE : LOW<br />

12<br />

SpO2<br />

PATIENT MODE : ADULT<br />

92<br />

NEXT CHANGE WAVE SEL PRINTER MORE<br />

Selects the next setting in the status window.<br />

Changes the currently selected display setting. (Pressing CHANGE<br />

at PATIENT MODE allows you to choose between Adult, Pediatric,<br />

and Neonatal in a Patient Mode window.)<br />

Allows access to the Printer Menu.<br />

Allows access to the Time/Day window.<br />

When the selected HR/PR source is no longer available, the current<br />

source is the active source with highest priority. The RR/BR source<br />

cannot be manually selected. It will always be CO 2 if CO 2 is active.<br />

Otherwise, it will be ECG/RESP.<br />

The user-selected HR/PR source is displayed along with the HR/PR<br />

source currently being used by the monitor.<br />

The selectable sweep speeds for HR/PR are 12.5, 25, and 50 mm/<br />

sec. The sweep speeds for RR/BR are 3.13, 6.25, and 12.5 mm/sec.<br />

Sets the Alarm Tone volume to HIGH, MEDIUM, or LOW.