20 Getting started <strong>Welch</strong> <strong>Allyn</strong> <strong>Propaq</strong> <strong>Encore</strong> <strong>Vital</strong> <strong>Signs</strong> <strong>Monitor</strong><br />
<strong>Monitor</strong> setup<br />
Setup Menu 1 is accessed by pressing the SETUP button on the Main Menu.<br />
STATSCALE<br />
ALARMS<br />
WAVE SEL<br />
TRENDS<br />
MORE<br />
STATSCALE<br />
ALARMS<br />
WAVE SEL<br />
TRENDS<br />
MORE<br />
NEXT<br />
CHANGE<br />
PRINTER<br />
MORE<br />
CURRENT<br />
SOURCE<br />
SELECTED<br />
SOURCE<br />
SWEEP (mm/s)<br />
ALARM TONE<br />
Automatically readjusts all waveform scales.<br />
Allows access to the Alarms menu.<br />
Allows you to turn on and off desired waveforms or NIBP numerics<br />
for display.<br />
Allows access to the Trend settings and display.<br />
Displays the next setup menu and the following status window:<br />
SETUP<br />
HR/PR RR/BR ( 85)<br />
CURRENT SOURCE : ECG MCO2 PA 35/ 18<br />
SELECTED SOURCE : ART --- MCO2 BR<br />
SWEEP (mm/s) : 12.5 3.13<br />
ALARM TONE : HIGH<br />
35<br />
mmHg<br />
HR/PR TONE : LOW<br />
12<br />
SpO2<br />
PATIENT MODE : ADULT<br />
92<br />
NEXT CHANGE WAVE SEL PRINTER MORE<br />
Selects the next setting in the status window.<br />
Changes the currently selected display setting. (Pressing CHANGE<br />
at PATIENT MODE allows you to choose between Adult, Pediatric,<br />
and Neonatal in a Patient Mode window.)<br />
Allows access to the Printer Menu.<br />
Allows access to the Time/Day window.<br />
When the selected HR/PR source is no longer available, the current<br />
source is the active source with highest priority. The RR/BR source<br />
cannot be manually selected. It will always be CO 2 if CO 2 is active.<br />
Otherwise, it will be ECG/RESP.<br />
The user-selected HR/PR source is displayed along with the HR/PR<br />
source currently being used by the monitor.<br />
The selectable sweep speeds for HR/PR are 12.5, 25, and 50 mm/<br />
sec. The sweep speeds for RR/BR are 3.13, 6.25, and 12.5 mm/sec.<br />
Sets the Alarm Tone volume to HIGH, MEDIUM, or LOW.