46 Patient monitoring <strong>Welch</strong> <strong>Allyn</strong> <strong>Propaq</strong> <strong>Encore</strong> <strong>Vital</strong> <strong>Signs</strong> <strong>Monitor</strong><br />

Manometer Bar.<br />

Systolic, diastolic, and<br />

mean values are indicated<br />

above the manometer as<br />

small triangles.<br />

Time of last measurement<br />

NIBP values and units of measure<br />

rotate in and out of this window<br />

for up to one hour (International<br />

English only)<br />

Measurement mode<br />

Time remaining in Turbocuf<br />

Mode or Auto Mode interval<br />

18:46:00 NIBP S 122 T1 102.3 F HR<br />

III<br />

80<br />

1cm<br />

mV<br />

ART<br />

MON<br />

109<br />

NIBP ADULT 18:45:28 mmHg ( 87)<br />

122 72<br />

MODE : TURBOCUF<br />

NIBP<br />

TIME : 4:21<br />

85 58 122<br />

SMARTCUF : ON<br />

85<br />

85 50 100 150 200<br />

58<br />

SpO2<br />

STOP<br />

Last measurement<br />

This symbol indicates the NIBP reading was taken in<br />

the presence of high motion artifact while<br />

monitoring ECG. Artifact can affect accuracy. To help<br />

reduce artifact, see page 45.<br />

99<br />

AUTO/MAN INTERVAL TURBOCUF SMARTCUF<br />

Alarm Limits Bell<br />

mmHg 122<br />

85 58<br />

NIBP values are<br />

displayed in one of<br />

these windows if<br />

vacant. After 16<br />

minutes, the display<br />

changes to that shown<br />

below for 44 minutes.<br />

NIBP 122/<br />

18:45 58<br />

( 85)<br />

START/STOP<br />

AUTO/MAN<br />

INTERVAL<br />

TURBOCUF<br />

SMARTCUF<br />

Starts and stops NIBP measurements. Any time the <strong>Propaq</strong> <strong>Encore</strong> is<br />

taking a noninvasive pressure measurement, the START button<br />

changes to STOP so you can stop the measurement in progress. This<br />

button initiates the same action as the START/STOP button at the left<br />

side of the screen. Pressing STOP will automatically vent the cuff.<br />

This button switches the mode between Automatic or Manual Mode.<br />

The Manual Mode is the default unless you change it by<br />

reprogramming your <strong>Propaq</strong>. Measurements can be taken at intervals<br />

of 1, 2, 3, 5, 10, 15, 30, and 60 minutes. Press START to initiate a<br />

measurement.<br />

Selects the interval at which NIBP measurements are automatically<br />

taken. The interval you select, ranging from one minute to 60 minutes,<br />

is shown on the display next to the word TIME.<br />

Automatically starts NIBP measurements and continues to take as<br />

many measurements as possible within five minutes.<br />

Enables or disables the Smartcuf motion artifact filter. NIBP<br />

measurements can still be taken when Smartcuf is off. Artifact may<br />

interfere with the accuracy of NIBP measurements with Smartcuf off.