46 Patient monitoring <strong>Welch</strong> <strong>Allyn</strong> <strong>Propaq</strong> <strong>Encore</strong> <strong>Vital</strong> <strong>Signs</strong> <strong>Monitor</strong><br />
Manometer Bar.<br />
Systolic, diastolic, and<br />
mean values are indicated<br />
above the manometer as<br />
small triangles.<br />
Time of last measurement<br />
NIBP values and units of measure<br />
rotate in and out of this window<br />
for up to one hour (International<br />
English only)<br />
Measurement mode<br />
Time remaining in Turbocuf<br />
Mode or Auto Mode interval<br />
18:46:00 NIBP S 122 T1 102.3 F HR<br />
III<br />
80<br />
1cm<br />
mV<br />
ART<br />
MON<br />
109<br />
NIBP ADULT 18:45:28 mmHg ( 87)<br />
122 72<br />
MODE : TURBOCUF<br />
NIBP<br />
TIME : 4:21<br />
85 58 122<br />
SMARTCUF : ON<br />
85<br />
85 50 100 150 200<br />
58<br />
SpO2<br />
STOP<br />
Last measurement<br />
This symbol indicates the NIBP reading was taken in<br />
the presence of high motion artifact while<br />
monitoring ECG. Artifact can affect accuracy. To help<br />
reduce artifact, see page 45.<br />
99<br />
AUTO/MAN INTERVAL TURBOCUF SMARTCUF<br />
Alarm Limits Bell<br />
mmHg 122<br />
85 58<br />
NIBP values are<br />
displayed in one of<br />
these windows if<br />
vacant. After 16<br />
minutes, the display<br />
changes to that shown<br />
below for 44 minutes.<br />
NIBP 122/<br />
18:45 58<br />
( 85)<br />
START/STOP<br />
AUTO/MAN<br />
INTERVAL<br />
TURBOCUF<br />
SMARTCUF<br />
Starts and stops NIBP measurements. Any time the <strong>Propaq</strong> <strong>Encore</strong> is<br />
taking a noninvasive pressure measurement, the START button<br />
changes to STOP so you can stop the measurement in progress. This<br />
button initiates the same action as the START/STOP button at the left<br />
side of the screen. Pressing STOP will automatically vent the cuff.<br />
This button switches the mode between Automatic or Manual Mode.<br />
The Manual Mode is the default unless you change it by<br />
reprogramming your <strong>Propaq</strong>. Measurements can be taken at intervals<br />
of 1, 2, 3, 5, 10, 15, 30, and 60 minutes. Press START to initiate a<br />
measurement.<br />
Selects the interval at which NIBP measurements are automatically<br />
taken. The interval you select, ranging from one minute to 60 minutes,<br />
is shown on the display next to the word TIME.<br />
Automatically starts NIBP measurements and continues to take as<br />
many measurements as possible within five minutes.<br />
Enables or disables the Smartcuf motion artifact filter. NIBP<br />
measurements can still be taken when Smartcuf is off. Artifact may<br />
interfere with the accuracy of NIBP measurements with Smartcuf off.