24 Getting started <strong>Welch</strong> <strong>Allyn</strong> <strong>Propaq</strong> <strong>Encore</strong> <strong>Vital</strong> <strong>Signs</strong> <strong>Monitor</strong><br />

Note<br />

The alarm for apnea cannot be turned off at any time.<br />

You can program the <strong>Propaq</strong> <strong>Encore</strong> to power up in any of the Factory patient modes or<br />

the Custom patient modes. You can also change the current patient mode during<br />

operation.<br />

Note<br />

Whenever you change the patient mode, the alarm limit settings automatically<br />

change to the settings for that mode.<br />

From the Main Menu, press SETUP, MORE, CHANGE, SETUP. The Mode Setup window<br />

appears:<br />

MODE SETUP BATTERY: 9.3 VOLTS ZEROED<br />

FACTORY: *ADULT PED NEO P2 NO ZERO<br />

MCO2<br />

CUSTOM :<br />

BR<br />

ADULT PED NEO<br />

* = PATIENT MODE ON POWERUP<br />

NEXT POWERUP* USE NOW SAVE<br />

38 12<br />

mmHg<br />

Sp02<br />

97<br />

PREV MENU<br />

The asterisk (*) indicates which patient mode is currently selected for powerup.<br />

NEXT Selects the next setting in the status window.<br />

POWERUP* Selects the highlighted patient mode (and its associated settings) as<br />

the powerup mode. The selected powerup mode is marked by an<br />

asterisk (*). (This does not change the current patient mode.)<br />

USE NOW Selects the highlighted patient mode (and its associated settings) as<br />

the current patient mode. (This does not affect the powerup mode.)<br />

SAVE Use this button to reprogram the settings of Custom patient mode:<br />

1. Make sure the patient mode you want to reprogram (ADULT, PED<br />

or NEO) is currently used. (To change patient modes, highlight the<br />

desired mode and press USE NOW, YES.)<br />

2. Exit the Mode Setup window, then use other menus and buttons to<br />

set the monitor settings and alarm limits as desired.<br />

3. Re-enter the Mode Setup window, highlight the desired Custom<br />

mode, and press SAVE, YES.