Empower Gas Chromatography - Waters
Empower Gas Chromatography - Waters
Empower Gas Chromatography - Waters
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
• Method parameters<br />
• Column type<br />
• Operating pressure<br />
• Carrier gas<br />
You might also be asked to provide the firmware versions of your GC instruments.<br />
To check the firmware version on the 5890, 6890, 6890 Plus or the 6890N GC, power on<br />
the instrument. The firmware version is displayed on the LCD.<br />
To check the firmware version on the injector towers and controllers:<br />
1. On the instrument keypad, press Options.<br />
2. Scroll down to Instrument Status and press Enter.<br />
3. Scroll down. The instruments are shown as follows:<br />
• Smplr:model&firmware<br />
• F inj:model&firmware<br />
• B inj:model&firmware<br />
4.1 <strong>Gas</strong> System Troubleshooting<br />
A poorly functioning gas system can cause problems in your <strong>Empower</strong> GC system. Before<br />
proceeding to Section 4.2, check these gas system components.<br />
Regulators<br />
To check the regulators, set the instrument for a constant pressure of 8 psi. Then use the<br />
manifold pressure (on the regulator, not the tank) to adjust the pressure to a constant flow<br />
at 8 psi.<br />
Tanks<br />
Ensure there is adequate pressure in the tanks. The lower the pressure in the tank, the<br />
more erratic the pressure and flow readings. As a rule, the tank should be replaced when<br />
its pressure falls below 100 psi. Follow your standard operating procedure for replacing the<br />
tank.<br />
4<br />
If you are using a split/splitless inlet, consider using the 6890 <strong>Gas</strong> Saver option, which<br />
reduces carrier gas consumption and allows you to reach pressures that may otherwise be<br />
difficult to achieve.<br />
Troubleshooting the GC System 44