Model 6100 and 6101 Ambulatory Infusion Pump
Model 6100 and 6101 Ambulatory Infusion Pump
Model 6100 and 6101 Ambulatory Infusion Pump
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Reference &<br />
Troubleshooting<br />
Section 6: Reference & Troubleshooting<br />
Message<br />
To continue, unlatch <strong>and</strong><br />
remove the Admin set or<br />
reservoir / Then reattach<br />
Upstream Occlusion /<br />
Press STOP to silence<br />
(<strong>Model</strong> <strong>6101</strong> only)<br />
Upstream Occlusion /<br />
Press STOP to silence /<br />
Press NEXT to restart<br />
(<strong>Model</strong> <strong>6101</strong> only)<br />
Wrong Cassette<br />
Corrective Action<br />
The cassette was not completely removed from the<br />
pump before it was reattached <strong>and</strong>, therefore, the<br />
pump’s sensors are not able to detect the cassette<br />
type. Remove the cassette <strong>and</strong> reattach it, then verify<br />
the cassette type in the pump’s display. If this alarm<br />
persists, remove the pump from use <strong>and</strong> contact<br />
Customer Service to return the pump for service.<br />
Fluid is not flowing from the fluid container to the<br />
pump. Check for a kink, a closed clamp, or air<br />
bubbles in the tubing between fluid container <strong>and</strong><br />
pump. Press ⁄ to stop the pump <strong>and</strong> silence the<br />
alarm for 2 minutes, then remove the obstruction <strong>and</strong><br />
press „ to restart the pump.<br />
Fluid is not flowing from the fluid container to the<br />
pump. Check for a kink, a closed clamp, or air<br />
bubbles in the tubing between fluid container <strong>and</strong><br />
pump. Press ⁄ to stop the pump <strong>and</strong> silence the<br />
alarm, then remove the obstruction <strong>and</strong> press „ to<br />
restart the pump.<br />
The pump detects the cassette is damaged, attached<br />
improperly, or incompatible with the pump. Close the<br />
tubing clamp. Make sure the cassette is attached<br />
properly. Then open the clamp <strong>and</strong> restart the pump.<br />
If the alarm persists, you may need to replace the<br />
cassette.<br />
97