Section 2: <strong>Pump</strong> Setup <strong>and</strong> Programming<br />
Setup &<br />
Programming<br />
TPN Delivery Method<br />
The TPN delivery method allows high volume delivery of solutions, with<br />
optional tapering. Delivery can be gradually increased or “tapered up” at the<br />
beginning of the infusion profile, or it can be gradually decreased, or “tapered<br />
down” at the end of the infusion profile. When the infusion profile is completed,<br />
the pump will beep 9 times. A Keep Vein Open (KVO) rate may be<br />
delivered at the end of the infusion profile, depending on the program you have<br />
entered.<br />
The following graph illustrates the TPN delivery method:<br />
Taper-Up<br />
<strong>Infusion</strong> Period<br />
Plateau Rate<br />
Taper-Down<br />
<strong>Infusion</strong> Volume<br />
KVO<br />
<strong>Infusion</strong> Profile<br />
TPN Delivery<br />
These are the programming screens for TPN Delivery. Complete descriptions of<br />
the screens follow.<br />
TPN Main Screen<br />
*** TPN 6230X ***<br />
STOPPED<br />
Press NEXT to advance<br />
Reservoir Volume<br />
Reservoir Volume<br />
| 1050 ml<br />
<br />
40