Section 2: <strong>Pump</strong> Setup <strong>and</strong> Programming<br />

to enter or verify the Continuous Rate <strong>and</strong> Dem<strong>and</strong> Dose. If the units are mg<br />

or mcg, you must also enter the Concentration. Changing the Units clears the<br />

amount Given <strong>and</strong> the Extended History.<br />

Setup &<br />

Programming<br />

Concentration<br />

If Units are mg or mcg, enter the concentration of drug in mg/ml or mcg/ml.<br />

When you enter a new Concentration, the pump requires you to enter a new<br />

Continuous Rate <strong>and</strong> Dem<strong>and</strong> Dose.<br />

Continuous Rate<br />

Enter the continuous rate of medication delivery (in mg/hr, ml/hr, or mcg/hr,<br />

depending on the Units). The maximum rate is 30 ml/hr or the mg or mcg<br />

equivalent. If the prescription does not call for a Continuous Rate, enter zero.<br />

NOTE: If you intend to run the pump in Lock Level 1 so the Continuous Rate<br />

can be varied, you should enter the maximum allowable rate while programming<br />

in Lock Level 0. After programming, you may then change to Lock Level<br />

1 <strong>and</strong> decrease the rate to its starting value. See Programming with Upper<br />

Limits, Adjusting Doses in LL1 at the end of this section<br />

Dem<strong>and</strong> Dose<br />

Enter the amount of drug to be delivered when the patient presses the Í key<br />

(or the Remote Dose Cord button, if attached). If the prescription does not call<br />

for a Dem<strong>and</strong> Dose, enter zero.<br />

NOTE: If you intend to run the pump in Lock Level 1 so the Dem<strong>and</strong> Dose<br />

can be varied, you should enter the maximum allowable dose while programming<br />

in Lock Level 0. After programming, you may then change to Lock Level<br />

1 <strong>and</strong> decrease the dose to its starting value. See Programming with Upper<br />

Limits, Adjusting Doses in LL1 at the end of this section.<br />

Dem<strong>and</strong> Dose Lockout<br />

If you programmed a Dem<strong>and</strong> Dose, enter the minimum amount of time that<br />

must elapse between the time one Dem<strong>and</strong> Dose starts <strong>and</strong> the time the next<br />

Dem<strong>and</strong> Dose starts. This lockout period is unaffected by removal of the<br />

battery or stopping of the pump.<br />

20