Chapter 4, <strong>Accu</strong>-<strong>Chek</strong> Report<br />

Evaluated data<br />

During creation of the individual analyses, the data read is<br />

checked by the <strong>Accu</strong>-<strong>Chek</strong> <strong>Smart</strong> <strong>Pix</strong> system. The following<br />

data is not included in the statistical analysis:<br />

• Test results outside the period selected<br />

• Test results saved without date and time<br />

• Tests with control solution<br />

• Invalid/deleted tests<br />

• Tests outside the measuring range<br />

(marked Hi/Lo)<br />

You will find some of this information summarized in the<br />

overview (e.g. number of Hi/Lo results) or in the logbook and<br />

the record list (e.g. results without a date or time, tests with<br />

control solution).<br />

Existing target range definitions are only adopted from the<br />

<strong>Accu</strong>-<strong>Chek</strong> Pocket Compass software. Apart from that, the<br />

settings made during configuration of the <strong>Accu</strong>-<strong>Chek</strong><br />

<strong>Smart</strong> <strong>Pix</strong> system apply.<br />

Please note the following when using the <strong>Accu</strong>-<strong>Chek</strong><br />

<strong>Smart</strong> <strong>Pix</strong> system with several meters:<br />

To ensure that the report belonging to a<br />

particular meter is displayed, compare the serial<br />

number on the meter with the serial number<br />

displayed in the report.<br />

4-4 <strong>Accu</strong>-<strong>Chek</strong> <strong>Smart</strong> <strong>Pix</strong> user's manual