<strong>CNS</strong> <strong>STANDARD</strong> <strong>OPERATING</strong> <strong>PROCEDURE</strong>LZR001, <strong>Two</strong> <strong>Photon</strong> Fluorescence Imaging SystemFigure 8: Beam routing of Ti:Sa on the optical table.Note that the laser beams are usually well aligned. Do not attempt torealign unless you are certain that the beams are off!6.2. Beam fine adjustment6.2.1. Make small adjustments to the steering mirrors to route the beam through the irisesinstalled in the beam path. The first mirror is the most likely mirror to have been bumpedby another user.SOP107Revision: 0.4Page 12 of 18 Effective Date: 04-13-2011