<strong>Advanced</strong> <strong>Instruments</strong> <strong>Inc</strong>.Mounting the <strong>Analyzer</strong>The standard <strong>GPR</strong>-<strong>2600</strong> is designed to be panel mounted and requires a cutout that accommodates the enclosure and 4mounting bolts. The design also lends itself to 19” rack mounting with an optional bezel or wall mount enclosures as illustratedbelow.Procedure:1. The <strong>GPR</strong>-<strong>2600</strong> is designed for panel mounting directly to any flat vertical surface, wall or bulkhead plate with theappropriate cut out and four ¼” diameter holes for insertion of the mounting studs located on the back side of the frontpanel.2. When mounting the analyzer position it approximately 5 feet off the floor for viewing purposes and allow sufficient room foraccess to the terminal connections at the rear of the enclosure.3. Note: The proximity of the analyzer to the sample point and use of optional sample conditioning components have animpact on sample lag time.Mounting <strong>GPR</strong>-<strong>2600</strong>:14