<strong>Advanced</strong> <strong>Instruments</strong> <strong>Inc</strong>.Mounting: The analyzer is approved for indoor use, outdoor use requires optional enclosures, consult factory. Mount asrecommended by the manufacturer.Gas Connections: Inlet and outlet vent gas lines for ppm analysis require 1/8” or ¼” stainless steel compression fittings; hardplastic tubing with a low permeability factor can be used percentage range measurements.Power: Supply power to the analyzer only as rated by the specification or markings on the analyzer enclosure. The wiring thatconnects the analyzer to the power source should be installed in accordance with recognized electrical standards. Ensure that isproperly grounded and meets the requirements for area classification. Never yank wiring to remove it from a terminalconnection. AC powered analog analyzers consume 5 watts, digital analyzers 50 watts without optional heaters. Optional 110Vand 220V heaters AC powered heaters consume an additional 100-150 watts; DC powered digital analyzers consume 30 watts,40 watts with the optional DC powered heater.8