11.07.2015 Views

GPR-2600 Series Oxygen Analyzer - Advanced Instruments Inc.

GPR-2600 Series Oxygen Analyzer - Advanced Instruments Inc.

GPR-2600 Series Oxygen Analyzer - Advanced Instruments Inc.

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Advanced</strong> <strong>Instruments</strong> <strong>Inc</strong>.The following display appears:Sample020.90 %Press UP or DOWN keys to change valuesSelect ENTER to save, ESC to returnAuto Range 0 to 1 %85⁰F 100Kpa 12/31/07 12:00:00Press the ENTER key to move the cursor (underscore) to the right to the digit to be changed.Press the UP or DOWN key until the desired number appears in digit field.Repeat as necessary to enter the numerical span value.Press the ENTER key to accept and save the span gas value.The system then initiates the SPAN CALIBRATION function and the following menu appears:CALIBRATIONIN PROGRESS . . .SPAN VALUEACTUAL O2 VALUEAfter the oxygen reading has stabilized, press ENTER to complete the Span Calibration.If the oxygen reading is outside +50% of the span gas value, the system displays the message “CALIBRATION FAILED” andreturns to the “Sample” mode.CALIBRATIONFAILED41