Vapor Generation Accessory VGAâ77 Operation manual
Vapor Generation Accessory VGAâ77 Operation manual
Vapor Generation Accessory VGAâ77 Operation manual
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
REDACIDSAMPLEACIDSAMPLEVGA–77REDUCTANTINERT GAS "B" TO SEPARATORINERT GAS "A" TOREACTION COILcell in cell holder atfront of AA module'JHVSFÁ Cell resting in module holder, when not in use*$1ÁNPEVMFÁIf you are using the ICP reagent module, follow the steps below toconnect the plasma gas/liquid separator outlet to the ICP instrument.✒ Note It is possible to connect the vapor generator to the ICP-OES or ICP-MSsystems while the plasma is on.1. Set up the sample introduction system as described in the ICPoperation <strong>manual</strong>.2. Use a sharp scalpel to neatly cut a length (appropriate to yourinstallation) of the black fluoro-elastomer tubing supplied.3. Fit the large black fluoro-elastomer tubing over the vapor outletof the plasma gas/liquid separator.4. Measure the minimum length of the supplied polyethylene tubingrequired to connect the free end of the black fluoro-elastomertubing to the sample inlet of the ICP nebulizer. Neatly cut therequired length and make the connection.✰✰✰ Hot Tip For mercury determinations using SnCl 2, the drain peristaltic pump tubingfrom the spectrometer should be pinched to prevent loss of gaseousmercury through the pumped drain.2-20 Publication date: May 2004