VGA–77'JHVSFÁ Plasma gas/liquid separator for ICP module2.2.12 Connecting a mercury trapYou should use a sharp scalpel to cut two suitable lengths of thesupplied 6 mm diameter clear PVC tubing, and connect them to thetwo cell ports. Connect the free ends to the outlet and then to asuitable mercury trap in accordance with local regulations governingthe discharge of mercury vapor.✰✰✰ Hot Tip Fitting is easier if the ends of the tubing are first softened in hot water.CautionBubble type collectors must not be used, because the VGA-77 is notdesigned to accommodate the excessive back pressure generated bysuch collectors. Using bubble type collectors may lead to vaporbeing forced through the drain of the gas/liquid separator andresults in an unreliable signal.Collectors of loosely packed metal foil may be suitable if theycomply with local lawsPublication date: May 2004 2-21