PERFORMANCE CHARACTERISTICSInput Power120 VAC ± 10% 220 VAC ± 10%Mains line frequency: 50 – 60 HzPower consumption: 240 VAFuses (two): 2.0 A (Slow Blow)Duty CycleUnder maximum power settings and rated load conditions (Pure Cut, 60 watt @ 500 Ω load), the generator issuitable for activation times of 10 seconds on, 30 seconds off for one hour.Dimensions and WeightMains line frequency: 50 – 60 HzPower consumption: 240 VAFuses (two): 1.0 A (Slow Blow)Width 26 cm (10.25 in.) Depth 11.4 cm (4.5 in.)Height 17.8 cm (7.0 in.) Weight < 15.4 kg (< 7 lbs)Operating ParametersAmbient temperature range 10° to 40° C (50° to 104° F)Relative humidityAtmospheric pressureWarm-up time30% to 75%, non-condensing700 hPa to 1060 hPaIf transported or stored at temperatures outside the operating temperature range,allow one hour for the generator to reach room temperature before use.Transport and StorageAmbient temperature range -34˚ to 65˚ C (-29˚ to 149˚ F)Relative humidity0% to 75%, non-condensingAtmospheric pressure500 hPa to 1060 hPaA-2 <strong>Bovie</strong> / <strong>Aaron</strong> <strong>Med</strong>ical