Audio VolumeThe audio levels stated below are for activation tones (bipolar, cut, and coag) and alarm tones (return electrode andsystem alarms) at a distance of one meter. Alarm tones meet the requirements for IEC 60601-2-2.Activation ToneVolume (adjustable)FrequencyDuration45 to 65 dBCut: 1 kHzBlend: 1 kHzCoagulation: 2 kHzFulguration: 2 kHzBipolar: 2 kHzContinuous while the generator is activatedAlarm ToneVolume (not adjustable)FrequencyDuration70 dB ± 5 dB2 kHz 1 ⁄2 seconds / 1 kHz 1 ⁄2 seconds2 sLow Frequency (50–60 Hz) Leakage CurrentEnclosure source current, ground open < 500 µASource current, patient leads, all outputsSink current at high line, all inputs < 50 µANormal polarity, intact ground: < 50 µANormal polarity, ground open: < 50 µAReverse polarity, ground open: < 50 µAHigh Frequency (RF) Leakage CurrentBipolar RF leakage currentMonopolar RF leakage current (additional tolerance)< 39 mA rms< 150 mA rms<strong>User</strong>’s Guide • <strong>Aaron</strong> <strong>950</strong>A-3