Aaron Bovie 950 User Manual - Med-E-Quip Locators
Aaron Bovie 950 User Manual - Med-E-Quip Locators
Aaron Bovie 950 User Manual - Med-E-Quip Locators
- No tags were found...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
OUTPUT CHARACTERISTICSMaximum Output for Bipolar and Monopolar ModesPower readouts agree with actual power into rated load to within 20% or 5 watts, whichever is greater.Mode Output Power Output Frequency Repetition Rate Vp-p maxCut 60 W @ 500 Ω 357 kHz ± 50 kHz N / A 1.5 KVBlend 60 W @ 800 Ω 357 kHz ± 50 kHz 30 kHz ± 5 kHz 2.0 KVCoagulation 60 W @ 1000 Ω 375 kHz ± 50 kHz 60 kHz ± 5 kHz 3.8 KVFulguration 35 W @ 1000 Ω 575 kHz ± 50 kHz 30 kHz ± 5 kHz 6.5 KVBipolar 30 W @ 200 Ω 520 kHz ± 50 kHz 19 kHz ± 5 kHz 2.0 KV<strong>User</strong>’s Guide • <strong>Aaron</strong> <strong>950</strong>A-5