OUTPUT CHARACTERISTICSMaximum Output for Bipolar and Monopolar ModesPower readouts agree with actual power into rated load to within 20% or 5 watts, whichever is greater.Mode Output Power Output Frequency Repetition Rate Vp-p maxCut 60 W @ 500 Ω 357 kHz ± 50 kHz N / A 1.5 KVBlend 60 W @ 800 Ω 357 kHz ± 50 kHz 30 kHz ± 5 kHz 2.0 KVCoagulation 60 W @ 1000 Ω 375 kHz ± 50 kHz 60 kHz ± 5 kHz 3.8 KVFulguration 35 W @ 1000 Ω 575 kHz ± 50 kHz 30 kHz ± 5 kHz 6.5 KVBipolar 30 W @ 200 Ω 520 kHz ± 50 kHz 19 kHz ± 5 kHz 2.0 KV<strong>User</strong>’s Guide • <strong>Aaron</strong> <strong>950</strong>A-5