12.07.2015 Views

900-C Conductivity, TDS, Resistivity, Ratio, Meter - TPS

900-C Conductivity, TDS, Resistivity, Ratio, Meter - TPS

900-C Conductivity, TDS, Resistivity, Ratio, Meter - TPS

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

5317.7 <strong>Conductivity</strong> Sensor Connections<strong>Conductivity</strong> sensor connections for <strong>TPS</strong> sensors are as follows…Pin No k=0.1 Sensor k=1.0 Sensor k=10 Sensor User k factor1 Cell Plate Cell Plate Cell Plate Cell Plate2 Link to Pin 6 Link to Pin 6 No Connection No Connection3 No Connection Link to Pin 6 No Connection No Connection4 Temp Sensor * Temp Sensor * Temp Sensor * No Connection5 Cell Plate Cell Plate Cell Plate Cell Plate6 Link to Pin 2.Temp Sensorcommon *Link to Pins 2and 3.Temp Sensorcommon *Temp Sensorcommon *No Connection* Temperature sensor connections to pins 4 and 6 are only applicable for combination conductivityand temperature sensors. Contact the factory for further details, if required. If in doubt do notmake any connections to these pins, and use a separate temperature sensor (<strong>TPS</strong> part number121245).k=0.1 Sensor k=1.0 Sensor k=10 Sensor3332 461 52 461 52 461 5Plugs shown above as viewed from the pin side of the plug.For non-<strong>TPS</strong> sensors, the links in the plugs may be fitted for matching nominal k factors.If a plug on a <strong>TPS</strong> sensor is replaced, be sure to install the links.