Chapter 4 Maintenance<strong>Series</strong> <strong>830</strong>/<strong>840</strong>/<strong>860</strong> Instruction ManualFigure 4-3. High Flow Transducer LFE CleaningHigh Flow Transducers:The high flow LFE is similar to the honeycomb element used in themedium flow body but larger in diameter. The high flow body consistsof four parts: inlet tube, inlet cap, main flow body <strong>and</strong> end cap.The inlet tube is only removed to inspect <strong>and</strong> replace the sealing O-ring between the inlet tube <strong>and</strong> inlet cap. To access the components:1. Remove the unit from the system.2. To remove the inlet screen, remove the four 1/4-28 socket head capscrews on the inlet side of the flow body <strong>and</strong> separate the inlet capfrom the main flow body.3. Inspect the inlet screen for damage <strong>and</strong> corrosion <strong>and</strong> replace if necessary.Light to medium particulate contamination can be cleaned by backwashing with a suitable solvent. Air dry thoroughly.4. Inspect the sealing O-ring for damage <strong>and</strong> replace if necessary.The inlet screen is mounted with the fine mesh side facing the inlet.5. To remove the LFE loosen <strong>and</strong> remove the four threaded rods holdingthe end cap to the main flow body. Separate the end cap from themain flow body <strong>and</strong> remove the LFE assembly. The LFE assemblyconsists of: 6-32 x 31/8 inch long screw, a #6 washer, two rangingwashers, honeycomb LFE, spacer, inlet filter <strong>and</strong> 6-32 nut.6. Inspect the honeycomb element for damage <strong>and</strong> replace if necessary.Replacement of the LFE or inlet screen requires transducer recalibration.`7. Re-assemble components. When the transducer is re-installed in the system,leak test the connection.8. Check transducer calibration.4-4 IM-83/84/86-H

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!