Sierra Series 830/840/860 Side-Trak™ and ... - Sierra Instruments
Sierra Series 830/840/860 Side-Trak™ and ... - Sierra Instruments
Sierra Series 830/840/860 Side-Trak™ and ... - Sierra Instruments
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Series</strong> <strong>830</strong>/<strong>840</strong>/<strong>860</strong> Instruction ManualChapter 5 TroubleshootingFlow Controllers with Motor-operated Rotary ValvesProblem Possible Cause SolutionValve “hunts” back <strong>and</strong> forthcontinuouslyValve rotates continuouslyValve does not rotatePressure differential too greatMotor coupling set screws looseValve motor failureLimit switch out of adjustment or brokenMicroprocessor failedContaminant blocking valveMotor fuse blownReduce pressureTighten set screwsReplace motorAdjust or replace switchReplace microprocessorRemove blockageReplace 1 amp fast blow fuse onmotor valve boardFlow reading is erratic Inlet filter screen clogged Clean or replace screenReads full scale with no flow orwith valve shutFlow does not match set pointwithin specificationsDefective sensorQ1 shortedU10 failedInsufficient pressure differentialReturn to factory for replacementReplace with same type partReplace with same type partIncrease pressure or reduce setpointIM-83/84/86-H 5-3