9 Maintenance and repair<br />
Maintenance work <strong>LUXOR</strong> S <strong>400</strong><br />
� Screw the hose [2] back on the nipple of the dry air filter [3].<br />
� The alarm message disappears and the alarm has been canceled.<br />
� Disconnect the dry air generator from the power supply.<br />
� Replace the lower front panel.<br />
9.5.4 Checking the resistance of the regeneration air heater<br />
� Prepare the dry air generator for maintenance (see “Maintenance and repair“ on Page 46).<br />
� No power supply is required to measure resistance.<br />
� Remove the back panel of the unit.<br />
� Measure the resistance on terminal strip –X13E1.<br />
� The resistance must be about 34 �.<br />
Power Voltage Resistance<br />
9 kW 230/<strong>400</strong> V approx. 34 �<br />
Fig. 9.6 – Measure resistance<br />
52 Edition: August 2005