30.03.2020 Views

Practical Gastrointestinal Endoscopy

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

THERAPEUTIC COLONOSCOPY 175

layer to steam. Because it is low voltage, however, cut current is

less able to traverse desiccated tissue and to heat deeply. ‘Autocut’

intelligent electrosurgical units automatically adjust power

output to match the resistance of the tissue being heated in order

to produce a predictable rate of transection.

Coagulating current has intermittent higher-voltage spikes with intervening

‘off periods’, which last for about 80% of the time (Fig.

7.9). The higher voltage allows a deeper spread of current flow

across desiccating tissue, whereas the off periods reduce (except

at high power settings) the tendency for gas ionization, sparking

and local tissue destruction.

Blended current combines both waveforms (Fig. 7.10), some units

providing the ability to select blends with relatively greater ‘cut’

than ‘coag’ characteristics. The differences between the various

makes of electrosurgical unit suggest that the output characteristics

are more complex than this brief summary suggests, some

appearing to provide more effective hemostasis than others.

When changing from one unit to another it is therefore essential

to be cautious, to start with low power settings. If possible, try out

the unit on a small lesion or the periphery of a larger one, rather

than entering the ‘big-time’ unrehearsed and then regretting it.

DESICCATION

Coagulating current—

intermittent high-voltage pulses

can pass desiccated tissue.

Fig. 7.10 Blended current combines

thecharacteristics of both

cutting and coagulating currents.

Current density

Tissue heats because of its high electrical resistance, typically

around 100 ohms, although resistance varies according to the

particular tissue (fat conducts poorly and so heats little); water

loss (desiccation) during heating increases resistance and the

drying tissue is also mechanically harder to transect. If electric

current is allowed to spread out and flow through a large area

of tissue, the overall resistance and heating effect falls (Fig. 7.11).

To obtain effective electrocoagulation, the flow of current must

be restricted through the smallest possible area of tissue—the

principle of ‘current density’ (Fig. 7.12). This principle is basic to

all forms of electrosurgery and explains why no noticeable heat

is generated at the broad area of skin contact with the patient

‘return plate’, whereas intense heat occurs in the closed snare

loop (Fig. 7.13). Even a relatively small area of contact between the

buttock or thigh and patient plate is adequate, and extra moisture

or electrode jelly is unnecessary at the power used for endoscopic

polypectomy.

The essential in polypectomy is to coagulate the core of the polyp stalk,

with its plexus of arteries and veins, before transection. Closing

the snare loop both stops the blood flow (‘coaptation’) and tends

to concentrate the current to flow through and heat-coagulate the

core (Fig. 7.14). The tightness of the loop is critical, since the area

through which the current is concentrated (current density) decreases

as the square of snare closure (πr 2 ), thus causing a square

law relationship between snare closure and increasing current

Fig. 7.11 Current flows more

easily through larger areas of

tissue resistance and so produces

little heat.

Fig. 7.12 Current density results

from constricting tissue and

greatly increases heating.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!