Capturing CO2 from ambient air - David Keith
Capturing CO2 from ambient air - David Keith
Capturing CO2 from ambient air - David Keith
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
In order to calculate <strong>CO2</strong> absorbed, outlet <strong>CO2</strong> concentration must be subtracted <strong>from</strong> background.<br />
This posed some challenges because background <strong>CO2</strong> sometimes varied dramatically, with spikes and<br />
drift on a time scale of 10 seconds to many minutes. These variations were often similar in magnitude to<br />
the absorption signal (4–50 ppm). Much of this appeared to be due to local variation <strong>from</strong> the ventilation<br />
system and <strong>from</strong> experimenters breathing near the inlet. We reasoned that connecting the inlet to the<br />
outside <strong>air</strong> would stabilize the <strong>CO2</strong> background. In fact outside concentration was at least as variable as<br />
inside concentration, perhaps due to signals <strong>from</strong> cars, plants, and drift <strong>from</strong> diurnal variation. The most<br />
stable background was achieved in trials 8–12 when the inlet was placed in a corner of the room and all<br />
experimenters stayed at the base of the tower, separated by about 3 m in height and 5 m in lateral distance<br />
<strong>from</strong> the inlet. Repeated measurements of the same conditions over time and across trials were largely<br />
consistent, so we were generally able to overcome background variation.<br />
Variation of absorption by NaOH concentration was measured by running separate trials, changing<br />
the working solution but holding other conditions constant. The main confound to this approach is that<br />
different concentrations have different densities and viscocities, so the drop distribution was different<br />
among the trials. The measured differences in <strong>CO2</strong> absorption are thus due to both changes in the reaction<br />
rate of <strong>CO2</strong> hydrolysis and to differing drop size.<br />
Variation of absorption by nozzle pressure was measured by changing the pressure (by adjustment of<br />
a needle valve) within the same trial. Using manufacturer data for changes of drop size and flow rate with<br />
pressure, the effect of drop size on absorption was inferred. The calculation is distorted by the use of 1.3<br />
M NaOH solution instead of water, which is what the manufacturer data apply to. However, since the<br />
solution is constant for the three pressures, the trend should be similar as for water.<br />
B.3 Data Analysis<br />
In the continuous mode trials, data were relatively easy to analyze when significant drift was not present.<br />
Figure B.8 shows the <strong>CO2</strong> concentration over a such a trial. The difference between the average background<br />
concentration and average steady-state outlet concentration gives the rate of <strong>CO2</strong> absorption. Some<br />
subjective judgment is required to choose the bounds of each period to average over. Periods were chosen<br />
to exclude transient regions between measuring states and regions which appear excessively noisy.<br />
We have samples of the solution taken periodically through some of the trials, including the one shown<br />
in Figure B.8, Trial 2. The CO 2−<br />
3 concentrations in those samples were measured with a Shimadzu Total<br />
Organic Carbon (TOC) analyzer after dilution and partial neutralization of the remaining NaOH. The<br />
results are shown in Figure B.9. If the rate of <strong>CO2</strong> uptake is taken as roughly constant, we would expect<br />
carbonate (CO 2−<br />
3 ) concentration in solution to increase linearly with operating time, assuming a constant<br />
solution volume. But the solution volume changes due to evaporation (and entrainment and solution stuck<br />
on surfaces, but we take these to be comparatively minor). Using the measured evaporation rate <strong>from</strong><br />
Trial 3 (which had similar conditions) to adjust the solution volume over time, we can make a prediction<br />
of CO 2−<br />
3 concentration based on the LiCOR measurements. This prediction is shown along with the TOC<br />
measurements in Figure B.9.<br />
For Trials 6–12, a more nuanced method of analysis was needed. The procedure in these trials involved<br />
81