WaveLink Series Differential Probe (13-25 GHz)<br />

Using Long Interconnect Extender (Provided with Dx10-, Dx20, and Dxx05-<br />

PT Kits) for Connecting Probe to EZ-Microtech Positioner Arm<br />

Attach the probe by first removing the screws holding the top plate to the V-shaped probe<br />

holder. Rest the Long Interconnect Extender in the V-shaped groove, and fasten the top plate to<br />

the holder, using the removed screws (as follows). Then, insert the Dx10, Dx20, or Dxx05-PT<br />

onto the Long Interconnect Extender.<br />

Figure 5-1. Positioner (Browser) Tip Mounted to EZ Probe Positioner and Long Interconnect Extender<br />

PLEASE NOTE THE FOLLOWING:<br />

� Do not overtighten the screws.<br />

� There is no fine adjust on the EZ Probe. It is easy to exceed the Z-Axis compliance rating<br />

of the PT. Use care when positioning the PT with the EZ Probe.<br />

Once the probe has been attached, loosen the knob on the EZ Positioner arm and position the<br />

probe close to the test point. Tighten the knob and use the joystick to fine position the probe.<br />

WL-HBW-OM-E RevC 38