1<br />
7. Heating and cooling technology<br />
Heating/Incubators<br />
Type Capacity Internal<br />
dimensions<br />
(W x D x H)<br />
756<br />
1<br />
Constant Climate Chamber Model HPP 108<br />
E & OE.<br />
www.wenk-labtec.com<br />
GENERAL CATALOGUE 2010/11<br />
Standard Features<br />
Memmert<br />
Ventilation and Control:<br />
- forced air, fan ventilation<br />
- adaptive, fuzzy-supported, multifunctional PID-controller, for stepless control of Peltier heating/cooling system -<br />
without refrigerant<br />
- autodiagnostic system with fault indication on temperature and humidity control<br />
- 2 x Class A, 4-wire circuit, Pt100 sensors, mutually monitoring and maintaining performance at the same<br />
temperature value<br />
- digital, 7-day, programme-timer with real time clock and precise minute setting<br />
- integrated timer for heating/cooling profiles of up to 40 ramps, each segment adjustable from 1 min. up to 999<br />
hours<br />
- digital display (LED) of all set parameters, such as temperature, day, time, humidity and set-up values - language<br />
chosen in setup<br />
- digital display of set /actual values: 0.1°C<br />
- GLP-compliant archiving of all relevant data, via 1024kB memory data logger<br />
- non-volatile memory<br />
- parallel printer interface (incl. real-time clock with date function) for all PCL3-compatible, ink-jet printers, providing<br />
GLP-compliant documentation<br />
- USB interface and Memmert "Celsius" software for programming and documentation<br />
- chip card control incl. one MEMoryCard XL with 32 kB (up to 40 ramps)<br />
- incl. works calibration certificate for +10°C and +37°C at 60% RH<br />
Humidifying and dehumidifying system:<br />
- active humidifying and dehumidifying from 10 to 90% RH with digital display (resolution : 0.5%, setting accuracy<br />
1% RH)<br />
- humidity supplied by distilled water from external tank using self-priming pump<br />
- humidification by hot steam generator<br />
- dehumidification by cold-trap using Peltier technology<br />
Multiple Overtemperature Protection<br />
- with audible and visual alarms in case of over-/under temperature and/or humidity, or empty water tank<br />
- independently working, digitally adjustable electronic overtemperature controller, TWW protection class 3.3<br />
- additional adjustable Auto-Safety-Function "ASF" for over- and undertemperature automatically following the set<br />
value at a preset tolerance range<br />
Textured Stainless Steel Housing:<br />
- fully insulated, stainless steel door with double locking and 4-point adjustment<br />
- inner glass door<br />
- rear panel: zinc-plated steel<br />
Interior:<br />
- easy-to-clean, hermetically welded, 1.4301 (ASTM 304) stainless steel interior, reinforced by deep-drawn ribbing,<br />
with 2 stainless steel wire shelves (HPP 749), 2 perforated stainless steel shelves (HPP 108)<br />
Heating- and cooling system:<br />
- energy-saving, Peltier heating-/cooling system integrated into the rear housing<br />
Temperature Range:<br />
- from +5°C up to +70°C<br />
- temperature variation in chamber at 10°C and 37°C: < ±0.4°C<br />
Voltage/Power Rating:<br />
- 230V (±10%) 50/60Hz<br />
Housing<br />
(W x D x H)<br />
litres mm mm kg W<br />
HPP 108 108<br />
560 x 400 x 480<br />
710 x 760* x 620<br />
66<br />
350<br />
HPP 749 749<br />
1040 x 600 x 1200 1190 x 825* x 1620 218<br />
1050<br />
* Depth without door handle, please add 38 mm<br />
E0 (X) Perforated stainless steel shelf<br />
Q1 IQ check list with works test data for unit as support for validation by customer<br />
Q3 Extra for FDA conformance software, "Celsius<br />
FDA-Edition"<br />
V1 Oven-linked authorisation card (user ID card)<br />
V2<br />
V4<br />
V3<br />
W4<br />
W6<br />
W7<br />
Interface RS485 (for cross linking of several devices) instead of USB<br />
MEMoryCard XL<br />
Temperature profile write/read unit for programming<br />
via PC, for writing to and reading from the chip card,<br />
up to 40 ramps<br />
Ethernet interface instead of RS 232 Software<br />
"Celsius- Ethernet-Edition" included<br />
RS232 instead of USB<br />
USB connection cable for computer interface<br />
Weight Rating PK Cat. No.<br />
1<br />
1<br />