GENERAL CATALOGUE 2010/11<br />

2<br />

Climatic Test Chamber CTC/Temperature Test chamber TTC<br />

Standard Equipment<br />

Ventilation and Control:<br />

- high-performance air fan in working chamber<br />

- depending on operation status automatically variable fan speed with manual adjustment from 10 to 100%<br />

- adaptive, fuzzy-supported multifunctional digital microprocessor PID-controller<br />

- autodiagnostic system with fault indication on temperature and humidity control<br />

- 2 x Class A, 4-wire circuit, Pt100 sensors, mutually monitoring and maintaining performance at the same<br />

temperature value<br />

- digital 7-day-programme-timer with real time clock, precise minute setting<br />

- integral timer for heating profiles of up to 40 ramps, each segment adjustable from 1 min. up to 999 hours<br />

- digital display (LED) of all set parameters, such as temperature, weekdays, time, humidity and set-up values -<br />

language to be chosen in setup<br />

- digital display of set values (resolution: 0.1°C below 99.9°C, 0.5°C above 100°C) and actual values (resolution:<br />

0.1°C) of temperature (LED)<br />

- 2 x 10 litre sumps acting as condensate collectors, on telescopic slide mounts<br />

- long-term documenting (ring store) of all relevant data, GLP-conforming as data logger - 1024 kB<br />

- programme stored on power failure<br />

- parallel printer interface (incl. real-time clock with date function) for all PCL3-compatible ink-jet printers for GLPconforming<br />

documentation<br />

- USB interface including Memmert "Celsius" software for programming and documentation<br />

- chip card control incl. one MEMoryCard XL with 32 kB (up to 40 ramps)<br />

- incl. works calibration certificate for -20°C and +160°C<br />

- incl. works calibration certificate for +30°C at 60% RH<br />

Heating concept:<br />

- high-performance ring heaters with optimised air circulation<br />

- door heating to avoid condensate<br />

Cooling Concept<br />

- twin compressor<br />

- speed adjustable condenser fan<br />

- chlorine-free refrigerant R404A<br />

Multiple Overtemperature Protection<br />

- with audible and visual alarm in case of over-/undertemperature and over-/under-humidity, open door and empty<br />

water tank<br />

- independently working, digitally adjustable electronic overtemperature controller, TWW protection class 3.3<br />

- additional adjustable Auto-Safety-Function "ASF" for over- and undertemperature automatically following the set<br />

value at a preset tolerance range<br />

- mechanical temperature limiter TB class 1 switching the heating off at approx.10°C above max. oven temperature<br />

Type Capacity Internal<br />

dimensions<br />

(W x D x H)<br />

CTC 256<br />

TTC 256<br />

Housing<br />

(W x D x H)<br />

litres mm mm kg W<br />

256<br />

256<br />

640 x 597 x 670<br />

640 x 597 x 670<br />

Accessories CTC 256/ TTC 256<br />

898 x 1100 x 1730<br />

898 x 1100 x 1730<br />

E & OE.<br />

www.wenk-labtec.com<br />

7. Heating and cooling technology<br />

Heating/Incubators<br />

Memmert<br />

Weight Rating PK Cat. No.<br />

340<br />

340<br />

7000<br />

7000<br />

2<br />

333<br />

1<br />

1<br />

9.868 010<br />

9.868 011<br />

Memmert<br />

Type Description PK Cat. No.<br />

BO<br />

E3<br />

W9<br />

Q1<br />

Q3<br />

V1<br />

V2<br />

V3<br />

V4<br />

Full-sight glass door<br />

Stainless steel grid<br />

External control and logging package<br />

IQ check list with works test data for unit as support for validation by customer<br />

Extra for FDA conformance software, "Celsius FDA-Edition"<br />

Oven-linked authorisation card (user ID card)<br />

RS485 interface (for cross linking of several devices) instead of USB<br />

Temperature profile<br />

MEMoryCard XL<br />

1<br />

1<br />

1<br />

1<br />

1<br />

1<br />

1<br />

1<br />

1<br />

9.868 015<br />

9.868 016<br />

9.868 017<br />

9.867 729<br />

9.867 730<br />

9.867 725<br />

9.867 713<br />

9.867 726<br />

9.867 727<br />

759