175A - HP Archive
175A - HP Archive
175A - HP Archive
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Section V<br />
Paragraphs 5-13 to 5-17<br />
i. Horizontal deflection should be 10 cm i0.5 cm.<br />
j. Check the two dc positions of HORIZONTAL<br />
DISPLAY in the same manner.<br />
k. Disconnect Calibration Generator.<br />
5-13. BANDWIDTH.<br />
a. Connect Signal Generator to HORIZ INPUT.<br />
b. Set:<br />
HORIZONTAL DISPLAY. . 0.1 VOLT/CM DC<br />
c. Set Signal Generator for 50 kc and adjust output<br />
amplitude for a horizontal deflection of 10 cm.<br />
d. Set frequency of Signal Generator to 500 kc.<br />
e. Horizontal deflection should be at least 7 cm.<br />
f. Disconnect Signal Generator.<br />
5-14. SWEEP GENERATOR.<br />
5-15. PRESET.<br />
a. Set:<br />
HORIZONTAL DISPLAY ......... X1<br />
SWEEP MODE ........... PRESET<br />
b. No sweep should be present.<br />
c. Connect CALIBRATOR 10 VOLT output to the<br />
vertical INPUT.<br />
d. A stable synchronized display should be present.<br />
e. Disconnect CALIBRATOR.<br />
5-16. TRIGGER POINT.<br />
a. Connect Oscillator to vertical INPUT.<br />
b. Adjust Oscillator output for 5 cmdeflection, and<br />
frequency to 1 kc.<br />
c. Rotate TRIGGER LEVEL to bothextremes. The<br />
starting point of the sweep should be adjustable to any<br />
point on the positive slope of the signal.<br />
d. Set:<br />
TRIGGER SLOPE ............. (-)<br />
e. Rotate TRIGGER LEVEL to bothextremes. The<br />
starting point of the sweep should be adjustable to any<br />
point on the negative slope of the signal.<br />
f. Disconnect Oscillator. Return TRIGGER LEVEL<br />
to 0 and TRIGGER SLOPE to +.<br />
5-17. TRIGGER SENSITIVITY.<br />
a. Connect Signal Generator to the verticalINPUT<br />
and TRIG INPUT.<br />
b. Set:<br />
SWEEP TIME .......... 1 pSEC/CM<br />
Vertical SENSITIVITY ... 0.5 VOLT/CM<br />
5-2<br />
Model <strong>175A</strong><br />
c. Set Signal Generator frequency to 1 mcandout-<br />
put amplitude for 1 cm of vertical deflection.<br />
d. Adjust TRIGGER LEVEL, if necessary, for a<br />
synchronized display and reduce the Signal Generator<br />
output amplitude until the sweep stops. The sweep<br />
should continue to trigger at signal amplitudes 0.2 cm<br />
or greater.<br />
e. Set:<br />
TRIGGER SOURCE ......... EXTAC<br />
f. Repeat steps c and d. The sweep should con-<br />
tinue to trigger at signal amplitudes of approximately<br />
0.5 cm or greater.<br />
g. Set:<br />
TRIGGER SOURCE ......... EXT DC<br />
h. Repeat steps c and d. The sweep should con-<br />
tinue to trigger at signal amplitudes of approximately<br />
0.5 cm or greater.<br />
i. Set frequency of Signal Generator to 50 mc and<br />
output amplitude for 3 cm of deflection.<br />
j. Set:<br />
SWEEP TIME ......... 0.1 pSEC/CM<br />
HORIZONTAL DISPLAY ......... XI0<br />
k. Repeat step d. The sweep should continue to<br />
trigger at signal amplitudes of approximately 1 cm<br />
or greater.<br />
m. Set:<br />
TRIGGER SOURCE ......... EXTAC<br />
n. Repeat step d. The sweep should continue to<br />
trigger at signal amplitudes of approximately 1 cm<br />
or greater.<br />
p. Set:<br />
TRIGGER SOURCE ........... INT<br />
q. Set Signal Generator output amplitude for 3 cm<br />
of deflection.<br />
r. Repeat step d. The sweep should continue to<br />
trigger at signal amplitudes of approximately 2 cm<br />
or greater.<br />
s. Disconnect Signal Generator.<br />
t. Connect a Model 10003A (AC-21M) Probe to<br />
vertical INPUT.<br />
u. Set:<br />
TRIGGER SOURCE ........... LINE<br />
Vertical SENSITIVITY .... 5 VOLTS/CM<br />
SWEEP TIME ......... 5MSEC/CM<br />
HORIZONTAL DISPLAY ......... X1<br />
TRIGGER LEVEL ............. 0<br />
v. Connect probe tip to power line. Donot connect<br />
probe ground lead.<br />
01526-1