Section V<br />
Paragraphs 5-47 to 5-50<br />
Figure 5-6. Attenuator Compensation<br />
(Top: undercompensated, Middle: compensated,<br />
Bottom: overcompensated)<br />
s. Disconnect 600-ohm Square- Wave Generator<br />
output from foam-covered wire and connect Generator<br />
to HORIZ INPUT. Adjust output amplitude for 8 cm<br />
deflection and frequency to 4 kc.<br />
t. Adjust Attenuator Compensation C203 for round<br />
dots with no tails (see figure 5-6).<br />
u. Disconnect Square- Wave Generator and jumper<br />
from vertical INPUT and sweep-generator printed-<br />
circuit board. Connect foam-covered wire to sweep-<br />
generator printed-circuit board.<br />
5-47. SWEEP GENERATOR.<br />
5-48. Refer to figure 5-3 for locationof adjustments.<br />
5-49. PRESET.<br />
a. Set:<br />
HORIZONTAL DISPLAY ......... XI<br />
SWEEP TIME ......... lOpSEC/CM<br />
b. Connect DC Voltmeter to pin 3 of V109. This<br />
voltage will be approximately -40 volts.<br />
c. Rotate SWEEP MODE slowly clockwise while<br />
watching Voltmeter. Meter pointer will jump when<br />
sweep generator free runs.<br />
. d. Note voltage at which sweep generator free runs<br />
(maximum negative reading obtained on Voltmeter).<br />
e. Set:<br />
SWEEP MODE. ........... PRESET<br />
f. Adjust Preset Adj R170 for meter reading of<br />
2.5 volts less negative than voltage noted in step d.<br />
5-10<br />
g. Disconnect Voltmeter.<br />
Model <strong>175A</strong><br />
5-50. TRIGGER SENSITIVITY AND SYMMETRY.<br />
a. Set:<br />
SWEEP MODE ........... PRESET<br />
TRIGGER SLOPE ............. (-)<br />
TRIGGER SOURCE ......... EXT AC<br />
HORIZONTAL DISPLAY ......... X1<br />
SWEEP TIME. ......... 1MSEC/CM<br />
Vertical SENSITIVITY. .. 0.05 VOLT/CM<br />
Vertical AC-DC ............. AC<br />
b. Connect Model 10003A (AC-21M) Probe from<br />
vertical INPUT to case of tunnel diode CR102 on<br />
sweep-generator circuit board (see figure 5-3).<br />
CAUTION: +110 VOLTS present at this point.<br />
c. Connect Calibration Generator to TRIG INPUT<br />
and adjust output for 0.05 volt rms.<br />
d. Mechanically center Trigger Symmetry R116,<br />
and turn Trigger Sensitivity R120 fully clockwise.<br />
e. Rotate TRIGGER LEVEL to obtain a free run-<br />
ning display approximately 0.5 volt in amplitude (see<br />
figure 5-7).<br />
Figure 5-7. Trigger Sensitivity<br />
(Top: Synchronized, Bottom: Free Running)<br />
f. Using TRIGGER LEVEL to maintain a presen-<br />
tation adjust Trigger Sensitivity R120 counterclock-<br />
wise until a stable, synchronized, symmetrical<br />
square wave approximately 0.35 volt in amplitude is<br />
obtained (see figure 5-7).<br />
g. Set Calibration Generator output to 0.03 volt rms.<br />
h. The trigger circuit should not trigger solidly.<br />
Counterclockwise adjustment of Trigger Sensitivity<br />
RH0 may be necessary to prevent solid triggering.<br />
Rotate TRIGGER LEVEL when checking for solid<br />
triggering.<br />
i. Set Calibration Generator output to 0.05 volt rms.<br />
j. The trigger circuit should again trigger solidly.<br />
Adjust TRIGGER LEVEL if necessary. Repeat steps<br />
f, g, h, and i until circuit triggers solidly with a<br />
0.05-volt rms input and does not trigger solidly with<br />
a 0.03-volt rms input.<br />
01526-1