Section V<br />

Paragraphs 5-47 to 5-50<br />

Figure 5-6. Attenuator Compensation<br />

(Top: undercompensated, Middle: compensated,<br />

Bottom: overcompensated)<br />

s. Disconnect 600-ohm Square- Wave Generator<br />

output from foam-covered wire and connect Generator<br />

to HORIZ INPUT. Adjust output amplitude for 8 cm<br />

deflection and frequency to 4 kc.<br />

t. Adjust Attenuator Compensation C203 for round<br />

dots with no tails (see figure 5-6).<br />

u. Disconnect Square- Wave Generator and jumper<br />

from vertical INPUT and sweep-generator printed-<br />

circuit board. Connect foam-covered wire to sweep-<br />

generator printed-circuit board.<br />

5-47. SWEEP GENERATOR.<br />

5-48. Refer to figure 5-3 for locationof adjustments.<br />

5-49. PRESET.<br />

a. Set:<br />

HORIZONTAL DISPLAY ......... XI<br />

SWEEP TIME ......... lOpSEC/CM<br />

b. Connect DC Voltmeter to pin 3 of V109. This<br />

voltage will be approximately -40 volts.<br />

c. Rotate SWEEP MODE slowly clockwise while<br />

watching Voltmeter. Meter pointer will jump when<br />

sweep generator free runs.<br />

. d. Note voltage at which sweep generator free runs<br />

(maximum negative reading obtained on Voltmeter).<br />

e. Set:<br />

SWEEP MODE. ........... PRESET<br />

f. Adjust Preset Adj R170 for meter reading of<br />

2.5 volts less negative than voltage noted in step d.<br />

5-10<br />

g. Disconnect Voltmeter.<br />

Model <strong>175A</strong><br />

5-50. TRIGGER SENSITIVITY AND SYMMETRY.<br />

a. Set:<br />

SWEEP MODE ........... PRESET<br />

TRIGGER SLOPE ............. (-)<br />

TRIGGER SOURCE ......... EXT AC<br />

HORIZONTAL DISPLAY ......... X1<br />

SWEEP TIME. ......... 1MSEC/CM<br />

Vertical SENSITIVITY. .. 0.05 VOLT/CM<br />

Vertical AC-DC ............. AC<br />

b. Connect Model 10003A (AC-21M) Probe from<br />

vertical INPUT to case of tunnel diode CR102 on<br />

sweep-generator circuit board (see figure 5-3).<br />

CAUTION: +110 VOLTS present at this point.<br />

c. Connect Calibration Generator to TRIG INPUT<br />

and adjust output for 0.05 volt rms.<br />

d. Mechanically center Trigger Symmetry R116,<br />

and turn Trigger Sensitivity R120 fully clockwise.<br />

e. Rotate TRIGGER LEVEL to obtain a free run-<br />

ning display approximately 0.5 volt in amplitude (see<br />

figure 5-7).<br />

Figure 5-7. Trigger Sensitivity<br />

(Top: Synchronized, Bottom: Free Running)<br />

f. Using TRIGGER LEVEL to maintain a presen-<br />

tation adjust Trigger Sensitivity R120 counterclock-<br />

wise until a stable, synchronized, symmetrical<br />

square wave approximately 0.35 volt in amplitude is<br />

obtained (see figure 5-7).<br />

g. Set Calibration Generator output to 0.03 volt rms.<br />

h. The trigger circuit should not trigger solidly.<br />

Counterclockwise adjustment of Trigger Sensitivity<br />

RH0 may be necessary to prevent solid triggering.<br />

Rotate TRIGGER LEVEL when checking for solid<br />

triggering.<br />

i. Set Calibration Generator output to 0.05 volt rms.<br />

j. The trigger circuit should again trigger solidly.<br />

Adjust TRIGGER LEVEL if necessary. Repeat steps<br />

f, g, h, and i until circuit triggers solidly with a<br />

0.05-volt rms input and does not trigger solidly with<br />

a 0.03-volt rms input.<br />

01526-1