Section V<br />

Table 5-9 (Cont'd)<br />

15. Trigger<br />

Sensitivity<br />

16. Trigger<br />

Symmetry<br />

Table 5-9. Condensed Adjustment Procedure (Cont'd)<br />

Ext Equip<br />

Test Procedure<br />

Adjust<br />

Required<br />

Calibration<br />

Generator<br />

Calibration<br />

Gene rator<br />

17. Sweep Length Square Wave<br />

Generator<br />

a. SWEEP MODE.. . . . PRESET<br />

TRIGGER SLOPE . . . . . . .<br />

(-)<br />

SWEEP TIME.. . 1 MSEC/CM<br />

SENSITIVITY. . . 0.05 V/CM<br />

b. Apply .05 vrms to TRIG<br />

INPUT (EXTAC)<br />

c. Connect probe to case of CR102<br />

d. Center Trigger Symmetry<br />

R116; Trigger Sensitivity RlZOfully<br />

cw (or ccw in some instruments).<br />

a. Complete step 15.<br />

b. Set TRIGGER SLOPE TO (+).<br />

c. Briefly ground grn/orn wire<br />

on SLOPE switch.<br />

Display 1 Mc squarewave at<br />

1 MSEC/CM.<br />

18. Sweep Time Sweep Time Markers Adjustment<br />

Time Mark<br />

Generator<br />

19. Vertical Gain Vertical Test<br />

Adapter; Calibra-<br />

tion Generator<br />

20. Scan Response Test Plug; Square<br />

Wave Generator;<br />

Test Oscilloscope<br />

21. Scan Balance Test Plug; Square<br />

Wave Generator;<br />

Test Oscilloscope<br />

Z2. Pulse Response Vertical Response<br />

Tester<br />

!3. High Frequency<br />

Response<br />

Vertical Response<br />

Tester<br />

!4. Calibrator Calibration<br />

Generator<br />

.I @EC/CM 10 Mc<br />

.2 @EC/CM 5 Mc<br />

.5 @EC/CM 1 psec<br />

1 @EC/CM 1 pec<br />

io @EC/CM 10 wec<br />

.I MSEC/CM 100 psec<br />

1 MSEC/CM 1 msec<br />

io MSEC/CM 10 msec<br />

.I SEC/CM 100 msec<br />

Apply lv p-p through Test Adapter<br />

to vertical amplifier<br />

Display 6 cm of 50 kc square wave,<br />

then install Test 'Plug (Figure 5-1),<br />

Trigger Test Oscilloscope on<br />

square wave.<br />

Complete step 20. Increase<br />

square wave frequency to 1 Mc.<br />

Observepositive pulse step at<br />

0.5 @EC/CM.<br />

Complete step 22. Increase<br />

SWEEP TIME to 0.1 @EC/CM.<br />

a. Generator output 1Ov p-p<br />

SENSITIVITY 1 V/CM.<br />

b. Reduce vertical deflection to<br />

exactly 6 cm with VERNIER.<br />

Model <strong>175A</strong><br />

a. Rotate TRIGGER LEVEL<br />

for free-running display.<br />

b. Trigger Sensitivity R120<br />

(with TRIGGER LEVEL) for<br />

stable, symmetrical square<br />

wave.<br />

c. If necessary, rotate R120<br />

ccw to prevent solid trig-<br />

gering on .03 vrms input.<br />

Trigger Symmetry R116<br />

for symmetrical square<br />

wave.<br />

Sweep Length R161 for 11<br />

cm of sweep<br />

C125<br />

C1016<br />

C1014<br />

c1012<br />

ClOlO<br />

R1005<br />

R1004<br />

R1003<br />

R1002<br />

Gain R45 for 5 cm deflection<br />

a. Scan Response C36 for<br />

best square wave observed<br />

at pin 15 of 5105;<br />

b. Scan Response C37 for<br />

best square wave at pin 13.<br />

Scan Balance C33 for bal-<br />

anced waveform observed at<br />

pin 10 of Test Plug. (Replace<br />

horizontal plug-in).<br />

a. Mid Freq. Response C22<br />

to drop leading corner 0.2 cm.<br />

b. Low Freq. Response C35<br />

for flat pulse top.<br />

c.Mid Freq. Response C22<br />

for best pulse response.<br />

Hi Freq. Response L7 and<br />

L8 for best high frequency<br />

corner. (Replace vertical<br />

plug-in).<br />

a. Connect CALIBRATOR<br />

10VOLToutput to INPUT.<br />

b. Cal Volt Adj R375 for<br />

exactly 6 cm deflection.<br />

5-20 01526-2