23.03.2013 Views

175A - HP Archive

175A - HP Archive

175A - HP Archive

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Section V<br />

Paragraphs 5-21 to 5-32<br />

5-21. CALIBRATOR.<br />

a. Connect the Calibration Generator to vertical<br />

INPUT and set output for 10 volts peak-to-peak.<br />

b. Set:<br />

SWEEP TIME .......... 2 MSEC/CM<br />

HORIZONTAL DISPLAY ......... X1<br />

Vertical SENSITIVITY ..... 1 VOLT/CM<br />

Vertical Sensitivity<br />

VERNIER .......... counterclockwise<br />

c. Rotate vertical sensitivity VERNIER clockwise<br />

for exactly 5 cm of vertical deflection.<br />

d. Disconnect Calibration Generator and connect<br />

CALIBRATOR 10 VOLT output to vertical INPUT.<br />

Vertical deflection should be 5 cm k0.05 cm if am-<br />

bient temperature is between 15°C and 35°C (50°F<br />

and 95°F). Vertical deflection should be 5 cm *0.15<br />

cm if ambient temperature is between 0°C and 15°C<br />

or 35°C and 55°C (32°F and 50°F or 95°F and<br />

130°F).<br />

5- 22. INTENSITY MODULATION.<br />

a. Connect the 600-ohm output of the Square Wave<br />

Generator to vertical INPUT and INTENSITY MOD-<br />

ULATION INPUT.<br />

b. Set 600-ohm output to maximum and frequency<br />

to 10 kc.<br />

c. Set:<br />

SWEEP TIME . ......... 20 pSEC/CM<br />

Vertical SENSITIVITY .... 10 VOLTS/CM<br />

INTENSITY MODULATION ........ EXT<br />

Vertical POLARITY ........... +UP<br />

d. The top of the square wave should beintensified<br />

with respect to the bottom.<br />

e. Disconnect Square Wave Generator. Return<br />

INTENSITY MODULATION to INT.<br />

5-23. SWEEP AND GATE OUTPUT<br />

(OPTIONAL FEATURE).<br />

a. Set SWEEP MODE to FREE RUN.<br />

b. Trigger Test Oscilloscope with gate output<br />

signal from Model <strong>175A</strong>.<br />

c. Observe SWEEP OUTPUT signal on Test Os-<br />

cilloscope. Signal should be a linear ramp of approx-<br />

imately -2.5 to +2.5 volts on all SWEEP TIME<br />

settings.<br />

d. Observe GATE OUTPUT signal on Test Oscil-<br />

loscope. Signal should be approximately 4 volts for<br />

the duration of sweep on all SWEEP TIME settings.<br />

e. Disconnect Test Oscilloscope.<br />

5-24. ACCESSORY PROBES.<br />

a. Set:<br />

SWEEP TIME ......... 0.5 MSEC/CM<br />

Vertical SENSITMTY ... 0.2 VOLTS/CM<br />

Model <strong>175A</strong><br />

b. Connect the DO A (AC- I) Probe from<br />

vertical INPUT to CALIBRATOR 10 VOLT output.<br />

c. Loosen locknut at rear of probe and rotate rear<br />

plastic flange with respect to cable. The waveform<br />

should change from one with overshoot to one with<br />

undershoot.<br />

d. Adjust probe for best square wave and tighten<br />

locknut. Note vertical amplitude of display.<br />

e. Disconnect probe from vertical INPUT and<br />

CALIBRATOR.<br />

f. Connect CALIBRATOR 1 VOLT output directly<br />

to vertical INPUT. The vertical amplitude of the<br />

display should be equal to that noted in step d.<br />

g. Disconnect CALIBRATOR from vertical INPUT.<br />

5-25. ADJUSTMENTS.<br />

5-26. This section covers internal adjustments of<br />

the Model <strong>175A</strong>. If difficulty is encountered in making<br />

any adjustment, refer to paragraph 5-67 for troubleshooting<br />

procedures. A condensed adjustment procedure<br />

is given in table 5-9.<br />

5-27. EQUIPMENT NEEDED FOR ADJUSTMENTS.<br />

5-28. In table 5-1, items 3 through 13 are the test<br />

instruments recommended to make the adjustments<br />

outlined in this section. Equipment with equivalent<br />

characteristics may be substituted. The special<br />

horizontal Test Plug shown in figure 5-1 is required<br />

to test the main vertical amplifier scanner outputs.<br />

5-29. LOCATION OF ADJUSTMENTS.<br />

5-30. Figures 5-2 and 5-3 show the location of tubes<br />

and adjustments in the Model <strong>175A</strong>. These diagrams<br />

are screened on the inside of the top and bottom<br />

covers of the instrument.<br />

5-31. PRELIMINARY SETTINGS.<br />

5-32. Install vertical and horizontal plug-in units<br />

in their compartments in the Model <strong>175A</strong>. If the<br />

vertical plug-in unit has more than one channel,<br />

select CHANNEL A and use channel A controls.<br />

a. Make the following settings on the panel of the<br />

vertical plug-in unit.<br />

AC-DC.. ................. AC<br />

VERNIER ................. CAL<br />

b. Depending on model number, make the follow-<br />

ing settings on the panel of horizontal plug-in unit:<br />

Model 1780A:<br />

SWEEP OCCURRENCE. .. .NORMAL<br />

Model 1781A/B:<br />

SWEEP SELECTOR .. .MAINSWEEP<br />

Model 1782A:<br />

FUNCTION ........... OFF<br />

Model 1783A:<br />

TIME MARKER. ........ OFF<br />

5-4 01526-2

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!