The Setup Screen displays the factory default settings the<br />
first time the dosimeter is used. Thereafter, stored settings<br />
are displayed.<br />
Measurement Mode<br />
NOTE: Exchange rate, criterion level and threshold level<br />
determine how noise dose is defined and computed, and<br />
will impact the dose results. The default settings correspond<br />
to NIOSH (1998) recommendations, which are more<br />
conservative (more protective) than the OSHA criteria.<br />
NIOSH (default)<br />
Exchange rate: 3 dB<br />
Criterion level: 85 dB<br />
Threshold level: 75 dB<br />
OSHA<br />
Exchange rate: 5 dB<br />
Criterion level: 90 dB<br />
Threshold level: 80 dB<br />
Custom<br />
Exchange rate: 3, 4 or 5 dB<br />
Criterion level: 70 to 95 dB, in 1-dB increments<br />
Threshold level: 65 to 90 dB, in 1-dB increments<br />
41