B800-00-880 Issue B<br />

Figure 12 - Controller status information<br />

Installation<br />

1. St<strong>and</strong>by speed increase button<br />

2. RS485/RS232 Switch<br />

3. Controller connector socket<br />

4. St<strong>and</strong>by speed reduce button<br />

5. Alarm LED<br />

6. Status LED<br />

7. Normal LED<br />

CAUTION<br />

When connecting the <strong>nEXT</strong> pump to a PC, remember that the 0 V pin on the RS232 connector may well be<br />

connected to earth through the PC. If this is the case ensure that the 0 V rail of the 24 - 48 V supply is not also<br />

connected to earth at some other point such as at the power supply. If you do not wish to have the 0 V rail of<br />

the 24 - 48 V supply connected to earth at the PC an opto-isolated interface to the PC should be used.<br />

The <strong>nEXT</strong> pump can connect directly to the RS485 or RS232 serial input on your control equipment or a PC as shown<br />

in Figure 13 <strong>and</strong> 14. In this configuration the PC is the serial link master <strong>and</strong> the <strong>nEXT</strong> pump is the slave. The distance<br />

over which the serial link will work is dependent on any difference in voltage between the 0 V at the sending <strong>and</strong><br />

receiving end. If the 0 V reference at the receiving end is within 0.3 V of the 0 V Control Reference pin on the <strong>nEXT</strong><br />

pump control connector then the serial link should be capable of operating at distances up to 6 m. An interface circuit<br />

external to the <strong>nEXT</strong> pump may be required for longer distances.<br />

The software in the <strong>nEXT</strong> pump is capable of operating with several pumps connected to a single serial link master.<br />

This is referred to as multi-drop mode. The RS485 option is recommended for multidrop mode. With the RS232 option<br />