B800-00-880 Issue B<br />

The pump will then accelerate to full operating speed. The green indicator LED will illuminate when the pump<br />

reaches Normal speed. (This is 80% of full rotational speed by default but you may have selected a different value to<br />

suit your application).<br />

4.5.3 St<strong>and</strong>by speed<br />

To run the <strong>nEXT</strong> pump at St<strong>and</strong>by Speed, send the following comm<strong>and</strong> over the serial communications link:<br />

Operation<br />

The reply you receive will be as follows:<br />

Comm<strong>and</strong> ! C 8 6 9 sp 1 cr<br />

Reply * C 8 6 9 sp r cr<br />

If the pump is currently below St<strong>and</strong>by Speed then it will accelerate until it reaches St<strong>and</strong>by Speed. If it is running<br />

faster than St<strong>and</strong>by Speed, it will decelerate until St<strong>and</strong>by Speed is reached.<br />

To return the pump to full speed, send the following comm<strong>and</strong>:<br />

The reply you receive will be as follows:<br />

Comm<strong>and</strong> ! C 8 6 9 sp 0 cr<br />

4.5.4 Stop the pump<br />

Reply * C 8 6 9 sp r cr<br />

To stop the <strong>nEXT</strong> pump, send the following comm<strong>and</strong> over the serial communications link:<br />

The reply you receive will be in the following format:<br />

Comm<strong>and</strong> ! C 8 5 2 sp 0 cr<br />

On successful receipt of the stop comm<strong>and</strong>, the pump rotor will decelerate to rest.<br />

4.5.5 Temperature readings<br />

Reply * C 8 5 2 sp r cr<br />

You can monitor the temperatures of both the pump motor <strong>and</strong> the internal electronics of the <strong>nEXT</strong> pump by sending<br />

the following query:<br />

Comm<strong>and</strong> ? V 8 5 9 cr<br />

The reply you receive will be as follows, where the first number is the motor temperature <strong>and</strong> the second number<br />

the internal Controller temperature, both measured in °C:<br />

Reply = V 8 5 9 sp d d d ; d d d cr<br />

4.5.6 Link parameter readings<br />

You can monitor the internal voltage, current <strong>and</strong> motor power of the <strong>nEXT</strong> pump by sending the following query:<br />

Comm<strong>and</strong> ? V 8 6 0 cr<br />

© <strong>Edwards</strong> Limited 2009. All rights reserved. Page 53<br />

<strong>Edwards</strong> <strong>and</strong> the <strong>Edwards</strong> logo are trademarks of <strong>Edwards</strong> Limited.