B800-00-880 Issue B<br />

6 Storage <strong>and</strong> disposal<br />

CAUTION<br />

Use of non-<strong>Edwards</strong> spares may result in reduced reliability <strong>and</strong> performance which will invalidate your<br />

warranty<br />

6.1 Storage<br />

Use the following procedure to store the pump.<br />

1. Place protective covers over the inlet, outlet, interstage (for the “i” <strong>nEXT</strong> only), purge, booster (for the “T”<br />

<strong>nEXT</strong> only) <strong>and</strong> vent ports.<br />

2. Place the pump in its packing materials. For fastest pump-down when the pump is put back into service, seal the<br />

pump inside a plastic bag together with a suitable desiccant.<br />

Storage <strong>and</strong> disposal<br />

3. Store the pump in cool, dry conditions until required for use. Refer to Table 14 for recommended storage<br />

environment. When required, prepare <strong>and</strong> install the pump as described in Section 3.<br />

4. Keep the pump upright at all times to prevent the drainage of oil from the bearing reservoir.<br />

5. Avoid long-term storage if possible. When long-term storage is necessary, the pump should be set up <strong>and</strong> run for<br />

at least eight hours every six months.<br />

6.2 Disposal<br />

WARNING<br />

In the unlikely event of a failure of the pump rotor, use appropriate personal protective equipment<br />

when h<strong>and</strong>ling <strong>and</strong> disposing of the pump <strong>and</strong> ensure that all pump inlets <strong>and</strong> outlets are capped<br />

off before disposal.<br />

Dispose of the <strong>nEXT</strong> pump <strong>and</strong> any components <strong>and</strong> accessories safely in accordance with all local <strong>and</strong> national safety<br />

<strong>and</strong> environmental requirements.<br />

Particular care must be taken with any components that have been contaminated with dangerous process substances.<br />

Take appropriate action to avoid inhalation of any particles that may be present in the pump.<br />

Do not incinerate the pump. The pump contains phenolic <strong>and</strong> fluorosilicone materials that can decompose to very<br />

dangerous substances when heated to high temperatures.<br />