B800-00-880 Issue B<br />

Operation<br />

4.1.6 Timer setting <strong>and</strong> options<br />

Refer to Section 1.4.3 for a full description of Timer functionality.<br />

The timeout period is a user-configurable option <strong>and</strong> can be set to any value from 1 to 30 minutes. When you receive<br />

the pump, it will be configured with a default timeout period of 8 minutes.<br />

To change the Timer setting, send the following comm<strong>and</strong> (where the characters 'd' represent the timeout period in<br />

minutes):<br />

The reply you receive will be as follows:<br />

Comm<strong>and</strong> ! S 8 5 4 sp d d cr<br />

Reply * S 8 5 4 sp r cr<br />

The Timer setting is now stored in memory within the <strong>nEXT</strong> pump.<br />

To check what timeout period is set, you can send a query as follows:<br />

The reply you receive will be as follows:<br />

Comm<strong>and</strong> ? S 8 5 4 cr<br />

Reply = S 8 5 4 sp d d cr<br />

The Timer is permanently enabled during ramp-up, however it is optional whether you have it enabled at other times.<br />

When you receive the pump, it will have the Timer enabled by default.<br />

If you wish to disable the Timer, send the following serial comm<strong>and</strong>:<br />

The reply you receive will be as follows:<br />

Comm<strong>and</strong> ! S 8 7 0 sp 0 cr<br />

Reply * S 8 7 0 sp r cr<br />

The state of the Timer option is stored in memory within the <strong>nEXT</strong> pump.<br />

To enable the Timer again, send the following serial comm<strong>and</strong>:<br />

The reply you receive will be as follows:<br />

Comm<strong>and</strong> ! S 8 7 0 sp 1 cr<br />

Reply * S 8 7 0 sp r cr<br />

To check whether the Timer is enabled or disabled, send the following query:<br />

Comm<strong>and</strong> ? S 8 7 0 cr<br />

The reply you receive will be as follows (where d=0 means disabled <strong>and</strong> d=1 means enabled):<br />

Reply = S 8 7 0 sp d cr<br />

Page 48<br />

© <strong>Edwards</strong> Limited 2009. All rights reserved.<br />