B800-00-880 Issue B<br />

Installation<br />

3.7.6 Multi-drop operation<br />

Using multi-drop mode, a single computer system can communicate with more than one <strong>nEXT</strong> pump. Each <strong>nEXT</strong> pump<br />

must be assigned its own individual address before it can be fitted into a multi-drop system. The comm<strong>and</strong> to assign<br />

the multi-drop address is sent in st<strong>and</strong>ard <strong>nEXT</strong> message format (as detailed in Section 4.1.9).<br />

The message protocol in multi-drop mode is marginally different to that described for serial messages in single pump<br />

systems. The main differences in multi-drop message protocol are detailed below:<br />

• All multi-drop comm<strong>and</strong>s, queries or replies have the first character #.<br />

• All comm<strong>and</strong>s, queries <strong>and</strong> replies contain a header containing the address of the node that the message is<br />

to, followed by the address of the node that the message is from.<br />

• There is a delimiter character: (colon) which separates the two multi-drop addresses in the header.<br />

• The remainder of the message (comm<strong>and</strong>, query or reply) follows the same protocol as already described for<br />

single pump systems.<br />

• The wild card address 99 is very useful <strong>and</strong> means ‘any’ node.<br />

After a <strong>nEXT</strong> pump has been assigned a multi-drop address, it will ignore any messages in the format for single pumps.<br />

An individual <strong>nEXT</strong> pump will remain silent <strong>and</strong> ignore all comm<strong>and</strong> messages unless the multi-drop address matches<br />

its own address.<br />

3.8 Connection for Mixed Parallel <strong>and</strong> Serial Operation<br />

You can control the pump using Parallel Interface control inputs <strong>and</strong> at the same time monitor various pump<br />

parameters using the Serial Interface. Alternatively you can control the pump using comm<strong>and</strong>s sent over the Serial<br />

Interface while at the same time monitor the Normal signal <strong>and</strong> Analogue Output over the Parallel Interface.<br />

Figure 17 shows a schematic diagram of a system that would allow you to do this. This connection is a hybrid of the<br />

parallel <strong>and</strong> serial connection which are described in detail in Section 3.6 <strong>and</strong> 3.7 respectively. Many of the options<br />

described in those sections are available in mixed parallel <strong>and</strong> serial operation but note that whilst serial enable is<br />

active to enable the serial link, the parallel st<strong>and</strong>by <strong>and</strong> fail signals are not available. The multidrop connection<br />

shown in Section 3.7. can also be used with mixed parallel <strong>and</strong> serial operation.<br />

You cannot control the pump using both the Parallel <strong>and</strong> Serial Interfaces simultaneously. For example, if you start<br />

the pump by sending a Start comm<strong>and</strong> over the Serial Interface you cannot then stop the pump by using the Start /<br />

Stop switch on the Parallel Interface. The pump will ignore the state of the Start / Stop switch on the Parallel<br />

Interface. To stop the pump you must send a Serial Stop comm<strong>and</strong>. Only when the Serial Stop comm<strong>and</strong> has been<br />

received by the pump can any comm<strong>and</strong>s sent via the Parallel Interface be acted on.<br />

Similarly, if you Start the pump by using the Start switch on the Parallel Interface you cannot then stop the pump by<br />

sending a Stop comm<strong>and</strong> over the Serial Interface. The pump will ignore any Start or Stop comm<strong>and</strong>s received over<br />

the Serial Interface. To stop the pump, you must use the Parallel Stop switch. Only when the pump has been stopped<br />

using the Parallel Interface switch will any Start or Stop comm<strong>and</strong>s be accepted via the Serial Interface.<br />

3.9 Vent Valve Selection, Connection <strong>and</strong> Control<br />

CAUTION<br />

If the pump is vented when it is at full rotational speed <strong>and</strong> the rate of pressure rise is too high, the pump could<br />

be damaged <strong>and</strong> its life may be reduced. We therefore recommend that either the rate of pressure rise is limited<br />

(refer to Figure 7) or that the vent valve is only opened after the <strong>nEXT</strong> pump speed has fallen to 50% of full<br />

rotational speed.<br />

To maintain the cleanliness of your vacuum system, we recommend that, whenever you switch the pump off, you<br />

vent the pump (or vacuum system) when the speed of the <strong>nEXT</strong> pump is between full rotational speed <strong>and</strong> 50% of full<br />

Page 40<br />

© <strong>Edwards</strong> Limited 2009. All rights reserved.<br />